ISOMETER® isoMED527 insulation monitor for medical IT systems, AC 100…240 V, NFC and RS-485
The isoMED527 monitors insulation in medical IT systems to IEC 60364-7-710 and IEC 61557-8. It also measures leakage capacitance, impedance, transformer load and temperature. NFC configuration, RS-485 interface.
- Purpose
- Insulation monitoring of medical IT systems
- Supply voltage
- AC 100…240 V
- Frequency
- 50…60 Hz
- Insulation resistance measuring range
- 1 kΩ…1 MΩ
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The isoMED527 serves un-earthed power supplies (IT systems) in medical locations and meets IEC 60364-7-710 and IEC 61557-8. It measures insulation resistance, leakage capacitance, impedance and total hazard current, and it monitors transformer load and temperature.
Insulation resistance is measured over 1 kΩ…1 MΩ. The response value is off or 50…500 kΩ, with a response time of ≤ 5 s, and the maximum permissible leakage capacitance is 5 µF. When a threshold is exceeded, LEDs, a relay with two changeover contacts, or the RS-485 and NFC interfaces report it. The isoNAP measuring method is used.
Settings are made over NFC with the Connect app, at an NFC frequency of 13.56 MHz. RS-485 runs at up to 115.2 kbit/s using Smart Connect (BSC). For fault location the device works with the ISOSCAN® EDS551, including the EDS551-6. STW2, STW3 and STW4 cover transformer load monitoring and the CP305 serves as remote alarm indicator.
A basic version, the isoMED527-L-2, and an advanced version with Function Module A activated are offered. Supply is AC 100…240 V, 50…60 Hz. Operating temperature is −25…+55 °C, maximum operating height 2000 m, protection IP20 and weight ≤ 130 g. The enclosure is 36 mm wide, 109 mm deep and 93 mm high.
Mounting is on a DIN rail to IEC 60715. The device complies with IEC 61557-8, IEC 60364-7-710, IEC 61010-1 and IEC 61010-2-30 and carries CE and UKCA marks.
Technical specifications
| Purpose | Insulation monitoring of medical IT systems |
|---|---|
| Supply voltage | AC 100…240 V |
| Frequency | 50…60 Hz |
| Insulation resistance measuring range | 1 kΩ…1 MΩ |
| Response value | off, 50…500 kΩ |
| Response time | ≤ 5 s |
| Maximum system leakage capacitance | 5 µF |
| RS-485 baud rate | up to 115.2 kbit/s |
| NFC frequency | 13.56 MHz |
| Maximum operating height | 2000 m |
| Operating temperature | -25…+55 °C |
| Degree of protection | IP20 |
| Weight | ≤ 130 g |
| Enclosure width × depth × height | 36 × 109 × 93 mm |
Part numbers
Show all part numbers (2)
B81625521, B81625524
How it compares with similar models
| Parameter | ISOMETER® isoMED527 insulation monitor for medical IT systems, AC 100…240 V, NFC and RS-485 | ISOMETER® isoMED427 insulation monitor for medical IT systems, AC 70…264 V | ISOMETER® isoLR275 with AGH-LR insulation monitor for IT systems, AC 0…793 V / DC 0…1100 V | ISOMETER® IR425-D4 insulation monitor for AC/DC IT systems 0…300 V, response values 1…200 kΩ |
|---|---|---|---|---|
| Purpose | Insulation monitoring of medical IT systems | Insulation monitoring of medical IT systems | Insulation monitoring of unearthed IT systems | Insulation monitoring of AC/DC control circuits (IT systems) |
| Operating temperature | -25…+55 °C | — | -25…+65 °C | -25…+55 °C |
| Weight | ≤ 130 g | — | ≤ 510 g | approx. 150 g |
Request a price and lead time
ISOMETER® isoMED527 insulation monitor for medical IT systems, AC 100…240 V, NFC and RS-485. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
What is the insulation resistance measuring range?
1 kΩ…1 MΩ, with the response value off or at 50…500 kΩ.
How is the isoMED527 configured?
Over NFC with the Connect app, or over RS-485.
How does the advanced version differ from the basic one?
The advanced isoMED527-L-2 has Function Module A activated; the basic device does not.