International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

ISOMETER® IR427 with MK7 insulation monitor for medical IT systems, AC 70…264 V

The IR427 with MK7 alarm indicator monitors insulation in medical IT systems to IEC 60364-7-710 and IEC 61557-8, and also load current and temperature. Two-module 36 mm enclosure, RS-485 interface.

Purpose
Insulation monitoring of medical IT systems
Supply voltage USA
AC 70…264 V, 42…460 Hz
Supply voltage MK7
DC 18…28 V
Maximum system leakage capacitance Ce
5 µF

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The IR427 is built for medical IT systems and follows IEC 60364-7-710 and IEC 61557-8. It measures insulation resistance and also monitors load current and temperature. Supply voltage USA is AC 70…264 V, 42…460 Hz, and the MK7 indicator panel runs on DC 18…28 V.

Alarms appear on LEDs, on an alarm relay and, over the bus, on the MK7 panel. Two response values are separately configurable, a multi-functional LC display shows the readings, and the device runs automatic and manual self-tests. An RS-485 interface is fitted.

The stated response values are Ran 50 kΩ for insulation resistance, Ialarm 7 A for current and 4 kΩ for temperature. The maximum permissible system leakage capacitance Ce is 5 µF, and a 6 A fuse is recommended.

The IR427 occupies a 36 mm two-module enclosure measuring 93 × 50 × 45 mm. The MK7 measures 80 × 47.5 × 74.5 mm and fits a standard flush-mounting box of 66 mm; the MK-Cavity-wall-box-60mm is offered for hollow walls.

The IR427 mounts on a DIN rail to IEC 60715 or with screws. Push-wire terminal versions carry part number B72075300 and screw-terminal versions B92075300. The measuring current transformer STW2 and the mounting frame XM420 are optional. The device meets IEC 61557-8, DIN VDE 0100-710 and EN 50110 and carries CE and UKCA marks, with REACH compliance.

Technical specifications

PurposeInsulation monitoring of medical IT systems
Supply voltage USAAC 70…264 V, 42…460 Hz
Supply voltage MK7DC 18…28 V
Maximum system leakage capacitance Ce5 µF
Response value Ran50 kΩ
Response value Ialarm7 A
Response value, temperature4 kΩ
Recommended fuse6 A
Enclosure width36 mm
Dimensions IR42793 × 50 × 45 mm
Dimensions MK780 × 47.5 × 74.5 mm
InterfaceRS-485

Part numbers

Show all part numbers (6)

B72075300, B92075300, B95100201, B942709, XM420B990994, B95100203

Request a price and lead time

ISOMETER® IR427 with MK7 insulation monitor for medical IT systems, AC 70…264 V. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

What does the IR427 monitor?

Insulation resistance, load current and temperature.

What system leakage capacitance is permitted?

The maximum permissible system leakage capacitance Ce is 5 µF.

Which connection versions are available?

Push-wire terminals under part number B72075300 and screw-type terminals under B92075300.

Similar equipment from other manufacturers

LET’S CONNECT

Let’s start with your challenge.

Complete a short brief. We will prepare an email you can send from your own email app.

Call us044 339-50-2009:00–18:00
USK ENERGY / SEARCH

Search

Solutions, specialist directions and guides