ISOMETER® isoMED427 insulation monitor for medical IT systems, AC 70…264 V
The isoMED427 monitors insulation in medical IT systems to IEC 60364-7-710 and watches transformer load and temperature. Versions marked P include a locating current injector. RS-485 interface (BMS).
- Purpose
- Insulation monitoring of medical IT systems
- Supply voltage
- AC 70…264 V, 47…63 Hz
- Maximum system leakage capacitance
- 5 µF
- Rated impulse voltage
- 4 kV
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The isoMED427 monitors insulation in medical IT systems to IEC 60364-7-710 and is supplied with AC 70…264 V, 47…63 Hz. Besides insulation resistance, it watches the load and temperature of the IT system transformer. Response values for insulation and for load current are adjustable.
Three versions are offered. The isoMED427-2 has no internal locating current injector. The isoMED427P-2 has one and holds a Lloyds Register approval. The isoMED427P-PT has the injector as well as temperature monitoring with a PT100 sensor. The injector switches on once an insulation fault is detected and works together with EDS series devices to find the fault.
Alarms and measured values go out over the RS-485 interface (BMS), and the BMS bus must be terminated at both ends. The measuring current transformer STW2 is available for transformer load monitoring.
The maximum permissible system leakage capacitance is 5 µF, the rated impulse voltage 4 kV and the rated insulation voltage 250 V. Power consumption is ≤ 3 W and the internal DC resistance ≥ 240 kΩ. Protection is IP30 for built-in components and IP20 for terminals. Dimensions are 93 × 70.5 × 36 mm, two modules wide.
The device mounts on a DIN rail or with screws, using one mounting clip per device, and a mounting frame XM420 is available. Terminals are push-wire (part numbers B7...) or screw-type (B9...). It complies with IEC 61557-8, IEC 61557-9 and DIN VDE 0100-710.
Technical specifications
| Purpose | Insulation monitoring of medical IT systems |
|---|---|
| Supply voltage | AC 70…264 V, 47…63 Hz |
| Maximum system leakage capacitance | 5 µF |
| Rated impulse voltage | 4 kV |
| Rated insulation voltage | 250 V |
| Power consumption | ≤ 3 W |
| Internal DC resistance | ≥ 240 kΩ |
| Protection, built-in components | IP30 |
| Protection, terminals | IP20 |
| Dimensions | 93 × 70.5 × 36 mm |
| Interface | RS-485 (BMS) |
| Versions | isoMED427-2; isoMED427P-2; isoMED427P-PT |
Part numbers
Show all part numbers (6)
B72075306, B92075306, B72075301, B92075301, B72075307, B92075307
Request a price and lead time
ISOMETER® isoMED427 insulation monitor for medical IT systems, AC 70…264 V. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
How do the isoMED427-2, isoMED427P-2 and isoMED427P-PT differ?
The isoMED427-2 has no locating current injector; the isoMED427P-2 has one and a Lloyds Register approval; the isoMED427P-PT has the injector plus PT100 temperature monitoring.
What leakage capacitance is permitted?
The maximum permissible system leakage capacitance is 5 µF.
Which terminal types are available?
Push-wire terminals with part numbers B7... and screw-type terminals with part numbers B9...
Similar equipment from other manufacturers