ES710 Single-Phase Isolating Transformers, 3150…10000 VA, for Medical IT Systems
The ES710 single-phase isolating transformers are for building medical IT systems to IEC 60364-7-710. Rated power 3150…10000 VA, input AC 230 V, output AC 230/115 V, reinforced insulation, with an energy-efficient Green Line (GL) edition.
- Rated power
- 3150…10000 VA
- Rated frequency
- 50…60 Hz
- Rated input voltage
- AC 230 V
- Rated output voltage
- AC 230/115 V
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The ES710 series has reinforced insulation and complies with DIN EN IEC 61558-1 (VDE 0570-1) and DIN EN IEC 61558-2-15 (VDE 0570-2-15). It also meets IEC 60364-7-710 / DIN VDE 0100-710 for IT systems in medical locations. The windings are galvanically isolated, with an electrostatic screen between them whose lead goes to an insulated terminal for the equipotential bonding. The fixing angles are insulated from the core.
Rated power is 3150…10000 VA at 50…60 Hz, with AC 230 V input, AC 230/115 V output and vector group IiO. DIN VDE 0100-710 requires the rated power of the transformer to be at least 3.15 kVA and at most 8 kVA. Inrush current IE is below 12 x In, and below 5 x In in the GL version. Leakage current is up to 0.5 mA, no-load input current i0 is up to 3 % (GL up to 2 %), no-load output voltage U0 is up to 236 V and short-circuit voltage Uk is up to 3 %. Full-load temperature rise is 48…70 K and the noise level is below 35 dB (A).
Designs are standard, S (terminals on top), GL (Green Line, energy-efficient), LG (horizontal mounting) and SK2 (encapsulated, protection class II, larger dimensions). The transformers suit horizontal and vertical installation in dry locations, and complete resin impregnation protects against corrosion. A PTC thermistor on each leg serves temperature monitoring, and the built-in sensors follow DIN 44081 (120 °C). Sheet-steel enclosures to IP 23 are available for all standard types.
Protection is IP00 (open design), in protection class I or II (encapsulated version), temperature class ta40/B, UL insulation system class F, with screw terminals. Weight starts from 15 kg and dimensions run from 230 x 325 x 204 mm (LG series) to 410 x 560 x 380 mm (SK2 series). The VDE ENEC mark is listed for the types. In a medical IT system overload protection is not permitted and short-circuit protection is required, while temperature protection is done with monitoring devices.
Technical specifications
| Rated power | 3150…10000 VA |
|---|---|
| Rated frequency | 50…60 Hz |
| Rated input voltage | AC 230 V |
| Rated output voltage | AC 230/115 V |
| Inrush current IE | < 12 x In (GL version < 5 x In) |
| Leakage current | ≤ 0.5 mA |
| No-load input current i0 | ≤ 3 % (GL ≤ 2 %) |
| No-load output voltage U0 | ≤ 236 V |
| Short-circuit voltage Uk | ≤ 3 % |
| Ambient temperature | ≤ 40 °C |
| Full-load temperature rise | ≤ 48…70 K |
| Noise level | < 35 dB (A) |
| Temperature class | ta40/B |
| UL insulation system | class F |
| Vector group | IiO |
| Degree of protection | IP00 (open design), IP23 (with enclosure) |
| Protection class | I, II (encapsulated SK2 version) |
| Connections | screw terminals |
| Weight | from 15 kg |
| Dimensions | 230 x 325 x 204 mm (LG) to 410 x 560 x 380 mm (SK2) |
Part numbers
Show all part numbers (6)
B92090001, B92090002, B92090003, B92090004, B92090005, B92090006 and others per the ordering table
How it compares with similar models
| Parameter | ES710 Single-Phase Isolating Transformers, 3150…10000 VA, for Medical IT Systems | DS0107 Three-Phase Isolating Transformers with Reinforced Insulation for Three-Phase Loads | ESL0107 Single-Phase Isolating Transformers, 120…1000 VA, for Operating Theatre Lights | WS50x80S…WS80x160S split-core current transformers for residual currents of 10 mA…100 A |
|---|---|---|---|---|
| Rated frequency | 50…60 Hz | 50…60 Hz | 50…60 Hz | 50…400 Hz |
| Vector group | IiO | Yyn O | IiO | — |
| Degree of protection | IP00 (open design), IP23 (with enclosure) | IP00 (open design), IP23 (with enclosure) | IP00 (open construction) | IP40 (internal components), IP20 (terminals) |
| Weight | from 15 kg | 16…155 kg | 0.5…5.8 kg | 900 g (WS50x80S), 1050 g (WS80x80S), 1250 g (WS80x120S), 2550 g (WS80x160S) |
Request a price and lead time
ES710 Single-Phase Isolating Transformers, 3150…10000 VA, for Medical IT Systems. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
What power rating is allowed for a medical IT system?
DIN VDE 0100-710 requires the rated power of the transformer to be at least 3.15 kVA and at most 8 kVA.
How does the GL version differ from the standard one?
The GL edition is energy-efficient: inrush current is below 5 x In against below 12 x In, and no-load input current i0 is up to 2 % against up to 3 %.
What protection is needed in a medical IT system?
Overload protection is not permitted and short-circuit protection is required; suitable fuses are selected from the technical data table.