Ta3000 gas analyser, isothermal chromatograph for trace impurities in gases, from 10 ppb

The ta3000 isothermal gas chromatograph measures impurities in pure and process gases from parts per million down to low ppb levels. It comes in two models, the ta3000R with a reduction gas detector and the ta3000F with a flame ionisation detector.

Measurement principle
isothermal gas chromatography
Detectors
RGD (ta3000R), FID (ta3000F)
Detection limit CO, CO2, CH4, C2H4
10 ppb
Detection limit H2 (ta3000R)
50 ppb

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The ta3000 is used for continuous monitoring of trace impurities in bulk gases such as oxygen, nitrogen, argon, helium, hydrogen and air. Its applications include oil and gas processing, industrial gas production, LCD, OLED and semiconductor manufacturing, and environmental and research laboratories. Typical tasks are measuring H2 and CO in ambient air and in ultra-high-purity gases, ethylene and ethylene oxide in air, CO in ethylene, propylene and propane, dissolved H2 in water, and checking the performance of gas purifiers.

The measurement is based on isothermal gas chromatography. Columns separate the sample components and the detector measures each one separately, so the influence of the permanent gas matrix is small. The ta3000R has a single high-sensitivity reduction gas detector (RGD) for hydrogen, carbon monoxide and unsaturated hydrocarbons. The ta3000F has a flame ionisation detector (FID) for carbon dioxide, methane and non-methane hydrocarbons. The ta3000F versions are designated by sample matrix: ta3000F-N2, ta3000F-Ar, ta3000F-He, ta3000F-O2 and ta3000F-H2, with a base ta3000F for air.

Detection limits are 10 ppb for CO, C2H4, CO2 and CH4, 50 ppb for H2 in the ta3000R, 30 ppb for ethylene oxide and 25 ppb for non-methane hydrocarbons. Display resolution is 0.1 ppb and resolution through the communication ports is 0.01 ppb, and the response to 99 % of the signal takes 2 to 10 minutes.

The instrument is designed for continuous operation. It mounts in a standard 19-inch industrial rack and can also be carried to site for surveys and one-off measurements. It has its own sample conditioning system and built-in data processing electronics. Together with a Sigma4000 multipoint stream switch, the analyser monitors several sampling points, because the built-in microprocessor controls the switch, stores the calibration parameters and processes the data.

Operating ambient temperature is 16 to 32 °C, and power is 100-120 or 200-240 VAC at 50/60 Hz, with a current draw of at least 5 A. Carrier gas enters at 4.8 to 6.2 bar through a UHP regulator, and the sample flow is 20 to 100 sccm. Dimensions are 426.7 x 177.8 x 673 mm and mass is 15.9 kg. The instrument has an RS232 port for a printer, a serial port for a PLC and an optional 4-20 mA output, and data is shown on a backlit LCD. It complies with the EU Electromagnetic Compatibility Directive.

Technical specifications

Measurement principleisothermal gas chromatography
DetectorsRGD (ta3000R), FID (ta3000F)
Detection limit CO, CO2, CH4, C2H410 ppb
Detection limit H2 (ta3000R)50 ppb
Detection limit ethylene oxide30 ppb
Detection limit non-methane hydrocarbons25 ppb
Range0–5 ppm
Resolution0.1 ppb (display), 0.01 ppb (ports)
Accuracy±1 x LDL or ±10 % of reading, whichever is greater
Response time2–10 min to 99 %
Ambient temperature16…32 °C
Power100–120 V or 200–240 V AC, 50/60 Hz
Carrier gas inlet pressure4.8–6.2 bar (70–90 psig)
Sample flow20–100 sccm
Dimensions (W x H x D)426.7 x 177.8 x 673 mm
Weight15.9 kg
InterfacesRS232, PLC serial port, 4–20 mA (option)

Part numbers

Show all part numbers (7)

ta3000R, ta3000F, ta3000F-N2, ta3000F-Ar, ta3000F-He, ta3000F-O2, ta3000F-H2

Request a price and lead time

Ta3000 gas analyser, isothermal chromatograph for trace impurities in gases, from 10 ppb. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

How do the ta3000R and ta3000F differ?

The ta3000R has a reduction gas detector and measures H2, CO and unsaturated hydrocarbons. The ta3000F, with a flame ionisation detector, measures CO2, CH4 and non-methane hydrocarbons.

Can one instrument monitor several sampling points?

Yes, in combination with a Sigma4000 multipoint stream switch, which the instrument's built-in microprocessor controls.

What are the carrier gas requirements?

An inlet pressure of 4.8 to 6.2 bar with ±2 % stability, supplied through a UHP regulator. The carrier gas temperature is 16 to 38 °C.