5100 TDLAS Tunable Diode Laser Gas Analyser for H2O, H2S, CO2, CH4 and O2
The 5100 TDLAS measures water vapour, hydrogen sulphide, carbon dioxide, methane or oxygen in process and flue gases using tunable diode laser absorption spectroscopy. A single absorption cell and an integrated sample conditioning system are housed in a compact IP65 enclosure.
- Measuring method
- tunable diode laser absorption spectroscopy (TDLAS), WMS
- Components measured
- H2O, H2S, CO2, CH4, O2, CO
- Range H2O
- 5-2500 ppmv
- Range CO2
- 0-500 ppmv
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The 5100 is a continuous analyser for one component in a gas stream at hydrocarbon processing and natural gas facilities. Typical tasks include moisture in transmission gas and at the outlets of molecular sieve dryers, hydrogen sulphide in fuel and flare gases, CO2 removal during gas sweetening, oxygen in process gases, combustion control, custody transfer metering, storage inlets and outlets, ethylene production, refining and petrochemicals.
The laser source and detector sit outside the sample gas, so contamination in the stream does not reach the optics. The laser is a distributed feedback type with an InGaAs photodiode detector, and wavelength modulation with 1F and 2F detection is applied. A sealed reference cell holding a known concentration of the target gas locks the laser to the chosen absorption line. Any drift of the spectrum serves as a correction signal, so the analyser confirms in real time that it measures at the correct wavelength. Signal processing is fully digital, and interference from methane, ethane and propane is compensated.
Wavelengths depend on the application: 1854 nm for water vapour, with 2004 nm, 1575 nm and 760 nm for other components. The ranges are 5 to 2500 ppmv for H2O, 0 to 500 ppmv for CO2, 0 to 100 ppmv for H2S and 0 to 10% for O2. Accuracy is 2% of reading or ±4 ppmv, whichever is greater. The photometric response is under 2 seconds, and T90 is 20 seconds at 2 L/min.
Sample flow is 1 to 2 L/min at a cell pressure of 70 to 170 kPa absolute, and the ambient range is -20 to 50 °C, with installation altitudes up to 2000 m. The enclosure is 304 stainless steel, rated IP65 and NEMA 4X. The explosion-protected version is marked Ex db eb IIB + H2 T3 Gb for Zone 1, with an ambient range of -40 to +50 °C, and CSA and NEC versions are also available.
Power is 120/240 VAC at 47 to 63 Hz or 24 VDC, drawing 25 W without the heater and 105 W with it. Communication uses Fast Ethernet, RS-485 with Modbus RTU, a 4-20 mA analogue output and four dry relay contacts rated 30 VAC, 60 VDC or 100 VA resistive. The analyser weighs 25 kg and measures 434 x 917 x 211 mm.
Technical specifications
| Measuring method | tunable diode laser absorption spectroscopy (TDLAS), WMS |
|---|---|
| Components measured | H2O, H2S, CO2, CH4, O2, CO |
| Range H2O | 5-2500 ppmv |
| Range CO2 | 0-500 ppmv |
| Range H2S | 0-100 ppmv |
| Range O2 | 0-10% |
| Accuracy | 2% of reading or ±4 ppmv, whichever is greater |
| Response time | under 2 s photometric; T90 20 s at 2 L/min |
| Sample flow | 1-2 L/min |
| Cell pressure | 70-170 kPa absolute (10-25 psia) |
| Ambient temperature | -20...50 °C |
| Power supply | 120/240 VAC, 47-63 Hz or 24 VDC |
| Power consumption | 25 W without heater, 105 W with heater |
| Outputs | Fast Ethernet, RS-485 (Modbus RTU), 4-20 mA, 4 dry relay contacts |
| Enclosure | IP65, NEMA 4X, 304 stainless steel |
| Explosion protection | Ex db eb IIB + H2 T3 Gb |
| Dimensions | 434 x 917 x 211 mm |
| Weight | 25 kg |
Request a price and lead time
5100 TDLAS Tunable Diode Laser Gas Analyser for H2O, H2S, CO2, CH4 and O2. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Does the 5100 TDLAS need frequent calibration?
No. A sealed reference cell keeps the laser locked to the absorption line, and the same cell allows continuous on-line verification of the analyser in operation.
Which outputs connect the analyser to the control system?
Fast Ethernet, RS-485 with Modbus RTU, a 4-20 mA analogue output and four dry relay contacts rated 30 VAC, 60 VDC or 100 VA resistive.
Can the analyser be installed in a hazardous area?
Yes. The version marked Ex db eb IIB + H2 T3 Gb is intended for Zone 1, with an ambient range of -40 to +50 °C.