5100HD TDLAS extractive laser gas analyser for 1 to 4 components and two sample streams
The 5100HD TDLAS is an extractive diode laser gas analyser with one or two absorption cells, measuring two streams or several components at once. Two lasers share one optical path, so a second analyser is not needed at the process point.
- Measurement method
- Tunable diode laser absorption spectroscopy (TDLAS), extractive
- Number of components
- 1-4 depending on configuration
- Sampling points
- 1 or 2
- H2O range
- 0-2500 ppmv
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The 5100HD is designed for hydrocarbon processing and natural gas, including streams with a high dew point. It measures combinations of water vapour, methane, hydrogen sulphide, carbon dioxide, acetylene and oxygen. Depending on configuration, it resolves one to four components and serves one or two sampling points. Special versions are offered for flare gas, landfill gas, and CO2 and O2 monitoring in ethylene oxide production.
Measurement is non-contact: the sample never touches the light source or detector, so fouling and damage to optics from the stream are excluded, and the response to sudden process changes is not delayed as it can be with contact sensors. Two separate lasers can be fitted in one unit with a shared optical path through the sample. A reference cell locks each laser onto the required absorption line, and any spectral shift is corrected in real time. Working wavelengths are 1854 nm for H2O, 2004 nm for CO2, 763 nm for O2 and 1575 nm for H2S.
Typical ranges are 0-2500 ppmv for moisture, 25-100% for methane, 0-300 ppm to 0-100% for H2S, 0-5% and 0-25% for O2, and 0-50 ppm to 0-100% for CO2. Moisture accuracy is ±4 ppmv or 2% of reading, and methane accuracy is ±2% of range. The photometric response is under 2 seconds, and T90 is under 50 seconds including sample system purge at a flow rate of 2 litres per minute.
The 300-series stainless steel enclosure carries IP 65 and Type 4X protection, so it can stand on an open site or sit in a hazardous area with an integrated sample conditioning system. Ambient temperature runs from -20 to 50°C and humidity from 0 to 90% non-condensing. Gas phase samples only are accepted, up to 150°C at 70-170 kPa absolute. The NEC/CEC Class I Division 2 and ATEX/IECEx Zone 2 version measures 880 x 674 x 302 mm and weighs 68 kg, while the purged Zone 1 version is 1113 x 866 x 323 mm and 82 kg, needing at least 140 litres per minute of purge air.
Power is 120/240 VAC at 47-63 Hz or 24 VDC, drawing 25-45 W without heating and 425-445 W with heating. Outputs include Fast Ethernet with Modbus TCP/IP, RS485 with Modbus RTU, up to two isolated 4-20 mA outputs and four dry relay contacts. The laser diode has a mean time between failures of over eight years.
Technical specifications
| Measurement method | Tunable diode laser absorption spectroscopy (TDLAS), extractive |
|---|---|
| Number of components | 1-4 depending on configuration |
| Sampling points | 1 or 2 |
| H2O range | 0-2500 ppmv |
| CH4 range | 25-100% |
| H2S range | 0-300 ppm to 0-100% |
| O2 range | 0-5% and 0-25% |
| CO2 range | 0-50 ppm to 0-100% |
| H2O accuracy | ±4 ppmv or 2% of reading |
| Response time | Under 2 s photometric; T90 under 50 s with purge at 2 L/min |
| Sample temperature | -20...+150°C, gas phase only |
| Sample pressure | 70-170 kPa absolute |
| Ambient temperature | -20...50°C |
| Power | 120/240 VAC 47-63 Hz or 24 VDC |
| Power consumption | 25-45 W without heater, 425-445 W with heater |
| Outputs | Fast Ethernet, Modbus TCP/IP, RS485 Modbus RTU, up to 2 x 4-20 mA, 4 relay contacts |
| Enclosure | IP 65/Type 4X, 300-series stainless steel |
| Dimensions and weight | 880 x 674 x 302 mm, 68 kg (Div 2/Zone 2); 1113 x 866 x 323 mm, 82 kg (Zone 1) |
Request a price and lead time
5100HD TDLAS extractive laser gas analyser for 1 to 4 components and two sample streams. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Can one 5100HD replace two analysers?
Yes, the unit can take two lasers on a shared optical path or two absorption cells, which allows up to four components or two streams to be measured.
What sample temperature can the analyser handle?
From -20 to +150°C, gas phase only, so the unit suits streams with a high dew point.
Is there a version for Zone 1?
Yes, the purged version for ATEX/IECEx Zone 1 measures 1113 x 866 x 323 mm and needs at least 140 litres per minute of purge air.
Similar equipment from other manufacturers