5100HD TDLAS extractive laser gas analyser for 1 to 4 components and two sample streams

The 5100HD TDLAS is an extractive diode laser gas analyser with one or two absorption cells, measuring two streams or several components at once. Two lasers share one optical path, so a second analyser is not needed at the process point.

Measurement method
Tunable diode laser absorption spectroscopy (TDLAS), extractive
Number of components
1-4 depending on configuration
Sampling points
1 or 2
H2O range
0-2500 ppmv

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The 5100HD is designed for hydrocarbon processing and natural gas, including streams with a high dew point. It measures combinations of water vapour, methane, hydrogen sulphide, carbon dioxide, acetylene and oxygen. Depending on configuration, it resolves one to four components and serves one or two sampling points. Special versions are offered for flare gas, landfill gas, and CO2 and O2 monitoring in ethylene oxide production.

Measurement is non-contact: the sample never touches the light source or detector, so fouling and damage to optics from the stream are excluded, and the response to sudden process changes is not delayed as it can be with contact sensors. Two separate lasers can be fitted in one unit with a shared optical path through the sample. A reference cell locks each laser onto the required absorption line, and any spectral shift is corrected in real time. Working wavelengths are 1854 nm for H2O, 2004 nm for CO2, 763 nm for O2 and 1575 nm for H2S.

Typical ranges are 0-2500 ppmv for moisture, 25-100% for methane, 0-300 ppm to 0-100% for H2S, 0-5% and 0-25% for O2, and 0-50 ppm to 0-100% for CO2. Moisture accuracy is ±4 ppmv or 2% of reading, and methane accuracy is ±2% of range. The photometric response is under 2 seconds, and T90 is under 50 seconds including sample system purge at a flow rate of 2 litres per minute.

The 300-series stainless steel enclosure carries IP 65 and Type 4X protection, so it can stand on an open site or sit in a hazardous area with an integrated sample conditioning system. Ambient temperature runs from -20 to 50°C and humidity from 0 to 90% non-condensing. Gas phase samples only are accepted, up to 150°C at 70-170 kPa absolute. The NEC/CEC Class I Division 2 and ATEX/IECEx Zone 2 version measures 880 x 674 x 302 mm and weighs 68 kg, while the purged Zone 1 version is 1113 x 866 x 323 mm and 82 kg, needing at least 140 litres per minute of purge air.

Power is 120/240 VAC at 47-63 Hz or 24 VDC, drawing 25-45 W without heating and 425-445 W with heating. Outputs include Fast Ethernet with Modbus TCP/IP, RS485 with Modbus RTU, up to two isolated 4-20 mA outputs and four dry relay contacts. The laser diode has a mean time between failures of over eight years.

Technical specifications

Measurement methodTunable diode laser absorption spectroscopy (TDLAS), extractive
Number of components1-4 depending on configuration
Sampling points1 or 2
H2O range0-2500 ppmv
CH4 range25-100%
H2S range0-300 ppm to 0-100%
O2 range0-5% and 0-25%
CO2 range0-50 ppm to 0-100%
H2O accuracy±4 ppmv or 2% of reading
Response timeUnder 2 s photometric; T90 under 50 s with purge at 2 L/min
Sample temperature-20...+150°C, gas phase only
Sample pressure70-170 kPa absolute
Ambient temperature-20...50°C
Power120/240 VAC 47-63 Hz or 24 VDC
Power consumption25-45 W without heater, 425-445 W with heater
OutputsFast Ethernet, Modbus TCP/IP, RS485 Modbus RTU, up to 2 x 4-20 mA, 4 relay contacts
EnclosureIP 65/Type 4X, 300-series stainless steel
Dimensions and weight880 x 674 x 302 mm, 68 kg (Div 2/Zone 2); 1113 x 866 x 323 mm, 82 kg (Zone 1)

Request a price and lead time

5100HD TDLAS extractive laser gas analyser for 1 to 4 components and two sample streams. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Can one 5100HD replace two analysers?

Yes, the unit can take two lasers on a shared optical path or two absorption cells, which allows up to four components or two streams to be measured.

What sample temperature can the analyser handle?

From -20 to +150°C, gas phase only, so the unit suits streams with a high dew point.

Is there a version for Zone 1?

Yes, the purged version for ATEX/IECEx Zone 1 measures 1113 x 866 x 323 mm and needs at least 140 litres per minute of purge air.

Similar equipment from other manufacturers

LET’S CONNECT

Let’s start with your challenge.

Complete a short brief. We will prepare an email you can send from your own email app.

Call us044 339-50-2009:00–18:00
USK ENERGY / SEARCH

Search

Solutions, specialist directions and guides