120HD Portable Hydrogen Analyser, Katharometer, 0 to 25% H2 in Air
The 120HD measures the hydrogen concentration in a gas sample and shows the hydrogen content in air as a percentage by volume on a digital display. It covers 0 to 25% H2 with an accuracy of ±1% H2 v/v and runs from a single 6 V alkaline battery.
- Measuring principle
- katharometer (thermal conductivity)
- Range
- 0 to 25% H2 in water-saturated air
- Accuracy
- ±4% of full scale (±1% H2 v/v)
- Repeatability
- ±2% of full scale (±0.5% H2 v/v)
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The instrument is used on ships, in turbine halls and during the charging of lead-acid batteries, places where hydrogen can collect in the air. It also suits research work and checks of hydrogen in air and in steam. Being fully portable and battery powered, it can be carried to the measurement point, and it leaves the factory calibrated.
Measurement follows the katharometer principle. The unit holds two parallel tubes, and each contains a gas and a heating coil. The composition of the sample is judged from how quickly heat is drawn from the coils into the gas. The coils sit in a bridge circuit, so a change in resistance caused by uneven cooling can be read directly. One channel holds air saturated with moisture, while the sample mixture passes through the other.
Hydrogen conducts heat far better than air, so the 120HD responds very strongly to it. Any gas whose thermal conductivity differs from that of air will also register, which matters when other gases are present in the sample. The test is non-destructive, so the sample can be used further after measurement. All wetted parts are stainless steel, and changes in ambient temperature and pressure do not affect the result.
The measuring range is 0 to 25% H2 in water-saturated air. Accuracy is ±4% of full scale (±1% H2 v/v), repeatability is ±2% of full scale (±0.5% H2 v/v), and the response time is under one minute for a 90% step change. The recommended sample flow is 50 ml/min, with a sample pressure of no more than 5 psi (34 kPa) gauge. The unit operates at ambient temperatures from 10 to 38 °C and relative humidity from 0 to 100%, and it passes the shock test MIL-S-901D (Navy) Grade B.
Power comes from one industrial alkaline lantern battery rated at 6 V. Battery life is at least 120 hours with the backlight on, and current draw is 210 mA with backlight or 170 mA without it. The unit weighs less than 15 lb with its battery fitted. Once the sample line is connected, check the joints for leaks.
Technical specifications
| Measuring principle | katharometer (thermal conductivity) |
|---|---|
| Range | 0 to 25% H2 in water-saturated air |
| Accuracy | ±4% of full scale (±1% H2 v/v) |
| Repeatability | ±2% of full scale (±0.5% H2 v/v) |
| Response time | under 1 minute for 90% step change |
| Ambient temperature | 10 to 38 °C |
| Relative humidity | 0 to 100% |
| Shock test | MIL-S-901D (Navy) Grade B |
| Sample flow rate | 50 ml/min |
| Sample pressure | no more than 5 psi (34 kPa) gauge |
| Power supply | 1 alkaline lantern battery, 6 V |
| Battery life | at least 120 hours with backlight |
| Current draw | 210 mA with backlight, 170 mA without |
| Weight | under 15 lb with battery |
Part numbers
Show all part numbers (1)
120HD
Request a price and lead time
120HD Portable Hydrogen Analyser, Katharometer, 0 to 25% H2 in Air. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which gases does the analyser respond to besides hydrogen?
Any gas whose thermal conductivity differs from that of air. The unit is most sensitive to hydrogen, whose thermal conductivity is far higher.
How long does the unit run on its battery?
At least 120 hours with the backlight on, from a single 6 V alkaline battery. Current draw is 210 mA with backlight and 170 mA without it.
Can the sample be used after analysis?
Yes. The method is non-destructive, so the sample remains usable after measurement.
Similar equipment from other manufacturers