International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

SKX9868 Flexible Type T Probe for Ultra-Low Air Temperatures

SKX9868 is a flexible Type T thermocouple probe for monitoring ultra-low air temperatures, including vaccine storage. Its range covers -196 to +100 °C, and it connects to the SRF614 transmitter through a 2-pin sub-miniature connector.

Sensor type
Type T thermocouple, class 3
Measuring range
-196 to +100 °C (-321 to +212 °F)
Accuracy
±1 °C (±2 °F) at -80 °C
Shaft length
200 mm

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The probe is intended for monitoring ultra-low temperatures in cold chambers and storage units where high measuring precision matters. The product literature names vaccine storage as a typical use. It works with the SRF614 transmitter, which forms part of the temperature monitoring system on site.

The sensing element is a class 3 Type T thermocouple built into a flexible wire assembly. The shaft measures 200 mm, while the PTFE insulated cable is 2 m long, bringing the overall probe length to 2.3 m. A 2-pin sub-miniature connector at the cable end lets the probe be connected to or removed from the transmitter without disturbing other joints.

The measuring range covers -196 to +100 °C, or -321 to +212 °F, and the accuracy is specified as ±1 °C (±2 °F) at -80 °C. For air temperature measurement, the shaft sits in the working volume of the chamber while the cable runs out to the transmitter, which is normally mounted outside. Over a long run, position the transmitter closer to the chamber to keep the cable short.

During installation, keep the cable out of door seals and away from sharp edges, which could damage the insulation. At scheduled checks of the monitoring system, compare the probe reading with a reference thermometer at the same point in the chamber. Before ordering, confirm which transmitter the site uses, where the shaft will be placed and how the 2 m cable will be routed. Production is certified to ISO 9001:2015.

Technical specifications

Sensor typeType T thermocouple, class 3
Measuring range-196 to +100 °C (-321 to +212 °F)
Accuracy±1 °C (±2 °F) at -80 °C
Shaft length200 mm
Cable length2 m
Overall length2.3 m
Cable insulationPTFE
Connector2-pin sub-miniature
Compatible transmitterSRF614

Part numbers

Show all part numbers (1)

SKX9868

Request a price and lead time

SKX9868 Flexible Type T Probe for Ultra-Low Air Temperatures. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Which transmitter does the SKX9868 probe work with?

It is compatible with the SRF614 transmitter and connects to it through a 2-pin sub-miniature connector.

What accuracy can be expected at very low temperatures?

The accuracy is specified as ±1 °C (±2 °F) at -80 °C, and the range spans -196 to +100 °C.

What are the shaft and cable dimensions?

Shaft length is 200 mm, the PTFE insulated cable is 2 m, and the overall probe length is 2.3 m.

Similar equipment from other manufacturers

LET’S CONNECT

Let’s start with your challenge.

Complete a short brief. We will prepare an email you can send from your own email app.

Call us044 339-50-2009:00–18:00
USK ENERGY / SEARCH

Search

Solutions, specialist directions and guides