PreMix 2000 analyser for excess O2 and excess fuel in premix burners
The stationary analyser measures excess oxygen or excess fuel in the air/fuel mixture before it reaches the burner. A zirconium oxide cell works in both oxidising and reducing mixtures, so one sensor shows which kind of mixture is present. The unit is intended for permanent installation in premix burner processes.
- Measurement principle
- zirconium oxide cell
- Range
- from 100% to 0.1% excess O2 and from 0.1% to 50% excess fuel
- Accuracy O2
- ± 2% of reading or ± 0.1% O2, whichever is greater
- Accuracy fuel
- ± 5% of reading or 0.25% excess fuel, whichever is greater
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The PreMix 2000 gives a continuous reading of the air/fuel ratio at premix burner stations, where a sample cannot be taken to a conventional flue gas analyser. Typical sites include glass forehearths and melting tanks, glass fibre apparatus, open flame brazing and soldering, flame treating, heat treatment of metals and ceramics, wire manufacturing, decarburising furnaces, ore roasting ovens and the primary stage of low NOx burners.
A small sample is drawn off, burnt, and the residual oxygen or excess fuel is measured. The sensing element is a zirconium oxide cell, and in oxidising mixtures its signal runs from 0 to about 130 mV. With excess fuel the signal climbs from about 550 mV to above 1000 mV, so one sensor shows whether the mixture is oxidising or reducing. Most of the flow goes through a bypass flow meter of up to 50 L/min, which shortens the response time and flushes stagnant volume from the inlet, while 0.5 L/min passes through the cell.
The sensor works with the Series 2000 microprocessor control unit. Its four-line, 20-character vacuum fluorescent display shows the readings, and the unit has two isolated current outputs that can be scaled freely. Either output can carry excess O2, excess fuel, a combined O2 and fuel scale, combustibles, the air/fuel ratio, cell temperature, or thermocouple or cell millivolts. The fuel type is picked from a list, or for a non-standard fuel the H/C ratio is entered, and the unit works out the calibration gas mix. Two independent alarms can each be set high or low.
The sensor is rated for ambient temperatures from -20 to 70 °C and has a UL Type 3R (IP14) enclosure. The control unit runs from -10 to 50 °C at 10% to 90% relative humidity without condensation. Its standard NEMA 4 (IP 56) enclosure suits wall or panel mounting, with general purpose wall, 19-inch rack and panel options, and a stainless steel NEMA 4X version. The sensor mounts on a panel, a wall or a pole. The portable CMFA-P 2000 holds the sensor and a Series 2000 control unit in a luggage-style case.
Technical specifications
| Measurement principle | zirconium oxide cell |
|---|---|
| Range | from 100% to 0.1% excess O2 and from 0.1% to 50% excess fuel |
| Accuracy O2 | ± 2% of reading or ± 0.1% O2, whichever is greater |
| Accuracy fuel | ± 5% of reading or 0.25% excess fuel, whichever is greater |
| Repeatability | ± 0.2% of reading |
| Sensor flow | 0.5 L/min |
| Bypass flow | up to 50 L/min |
| Sample pressure | not over 10 psig |
| Sensor ambient temperature | -20 to 70 °C |
| Sensor power | 115 VAC ± 10%, 1200 VA max; 230 VAC ± 10%, 2400 VA max |
| Sensor enclosure | UL Type 3R (IP14) |
| Analogue outputs | two isolated, 4-20 / 0-20 / 20-4 / 20-0 mA, load up to 1200 ohms |
| Alarms | two independent, high or low, contacts 0.5 A, 30 V, 10 VA |
| Communications | RS-485, addressable, two-way |
| Control unit ambient temperature | -10 to 50 °C |
| Control unit power | 115-230 VAC ± 10%, 47-63 Hz, 75 VA max |
| Control unit enclosure | NEMA 4 (IP 56); optional NEMA 4X (IP 56) |
| Sample inlet | 1/4 inch compression fitting |
Request a price and lead time
PreMix 2000 analyser for excess O2 and excess fuel in premix burners. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Does the analyser work in a reducing atmosphere?
Yes. It measures excess fuel in the range 0.1–50%, and in that mode the cell signal runs from about 550 mV to above 1000 mV.
Which signals does the control unit provide?
Two isolated current outputs of 4-20, 0-20, 20-4 or 20-0 mA into loads up to 1200 ohms, two alarm contacts rated 0.5 A at 30 V, and an RS-485 interface.
Is there a portable version?
Yes. The CMFA-P 2000 houses the sensor and the Series 2000 control unit in a luggage-style case, and it can measure both the mixture feeding a burner and flue gases.