CEM/O2 Extractive Oxygen Analyser with Zirconia Cell, 0.1 to 100% O2
The CEM/O2 is an extractive analyser for pure oxygen, built around a zirconium oxide cell, for continuous emissions monitoring systems on wet or dry gas. It covers 0.1 to 100% O2 with an accuracy of ±0.75% of reading or 0.05% O2, and its response time is under 4 seconds.
- Measuring principle
- ZrO2 cell (zirconium oxide)
- Measuring range
- 0.1–100% O2
- Accuracy
- ±0.75% of reading or 0.05% O2, whichever is greater
- Response time
- under 4 s at 2 scfh (from 2% to 20% O2)
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The CEM/O2 measures the content of pure oxygen in flue gases and process streams. The analyser works within continuous emissions monitoring (CEM) systems on wet or dry gas, where the sample is drawn off and conveyed to the point where the instrument sits.
The sample inlet feeds a direct sampling channel with a centrally placed convection tube that works on the draught principle of the stack. Part of the sample passes through the convection loop, where a zirconia (ZrO2) cell measures the oxygen. The inlet temperature can reach 204 °C (400 °F), the sample pressure is held to ±2 psig (0.14 kg/cm2), and the flow runs from 2 to 20 scfh (0.94 to 9.4 L/min). At 2 scfh the response time is under 4 seconds for a change from 2% to 20% O2, and drift stays below 0.1% of cell output per month.
The compact sensor housing is installed in CEM cabinets, and the design meets the requirements of 40 CFR Part 60 and Part 75. For sample streams carrying dust or condensate, the CEM O2/TM has a sampling design that is insensitive to flow, and it covers 1 ppm to 100% O2 with a response time under 10 seconds at 1.0 L/min. The CEM/Humox version measures moisture in flue gas using two sensors, one dry and one wet, and its control unit calculates moisture from the difference in O2 concentration across a range of 5% to 85%.
The analyser operates with the Series 2000 control unit, which has a four-line, 20-character vacuum fluorescent display, analogue outputs, alarms and extended system diagnostics, with optional auto-calibration. Two isolated current outputs can each be set to 4-20, 0-20, 20-4 or 20-0 mA, with a maximum load of 1200 ohms. The relay contacts are rated at 0.5 A, 30 V and 10 VA, and data are exchanged over two-way addressable RS-485.
The control unit is normally mounted in a 19-inch rack, with optional general-purpose wall or panel mounts and NEMA 4 and NEMA 4X (IP56) housings. The sensor is for indoor use only. Ambient temperature runs from -18 to 50 °C, relative humidity from 10% to 90% without condensation, with operation up to 2000 meters altitude. Supply is 115 or 230 V AC ±10% at 47 to 63 Hz. The instrument complies with the EMC directive 89/336/EEC and the low voltage directive 73/23/EEC, and holds UL listing for NEC Class I, Division 2 areas.
Technical specifications
| Measuring principle | ZrO2 cell (zirconium oxide) |
|---|---|
| Measuring range | 0.1–100% O2 |
| Accuracy | ±0.75% of reading or 0.05% O2, whichever is greater |
| Response time | under 4 s at 2 scfh (from 2% to 20% O2) |
| Drift | under 0.1% of cell output per month |
| Sample inlet temperature | up to 204 °C (400 °F) |
| Sample pressure | ±2 psig maximum (0.14 kg/cm2) |
| Sample flow | 2 to 20 scfh (0.94 to 9.4 L/min) |
| Analogue outputs | 2 isolated, 4–20 / 0–20 / 20–4 / 20–0 mA |
| Maximum load | 1200 ohms |
| Relay contacts | 0.5 A, 30 V, 10 VA, AC or DC |
| Interface | RS-485, two-way addressable |
| Display | vacuum fluorescent, 4 lines × 20 characters |
| Power supply | 115 or 230 V AC ±10%, 47–63 Hz |
| Ambient temperature | −18 to 50 °C |
| Relative humidity | 10–90%, non-condensing |
| Control unit housing | 19-inch rack; optional NEMA 4 and NEMA 4X (IP56) |
| Variants | CEM/O2, CEM O2/TM, CEM/Humox |
Request a price and lead time
CEM/O2 Extractive Oxygen Analyser with Zirconia Cell, 0.1 to 100% O2. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Does the CEM/O2 work on wet gas?
Yes. The analyser is used in CEM systems on both wet and dry gas, and it is designed for hot wet extractive systems.
How can flue gas moisture be measured?
With the CEM/Humox version, which has two CEM/O2 sensors, one dry and one wet. The control unit calculates moisture over a range of 5% to 85% from the difference in O2 concentration.
Which version suits samples with dust and condensate?
The CEM O2/TM, whose sampling design is insensitive to flow. It covers 1 ppm to 100% O2 with a response time under 10 seconds at 1.0 L/min.