5920 UHP Moisture Analyser for Ultra High Purity Gases, 0 to 150 ppbv
The 5920 measures moisture at the parts per billion level in ultra high purity gases. It uses a quartz crystal microbalance sensor with an accuracy of ±1 ppbv or ±10% of reading, and includes built-in online zero checking and gas type selection from the menu.
- Measured component
- H2O
- Principle
- Quartz crystal microbalance (QCM)
- Range
- 0–150 ppbv; trend indication to 1000 ppbv
- Detection limit
- 1 ppbv
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The 5920 is intended for critical trace moisture measurements: in high purity gas production, air separation plants, semiconductor and LCD/OLED display manufacturing, continuous monitoring and spot checks, including gas cylinders. It is also used for hydrogen in hydrogen cooled generators and for blanketing gases in steel heat treatment.
Measurement is made by a quartz crystal microbalance (QCM) sensor. The detection limit is about 1 ppbv and sensitivity is better than 1 ppbv. The calibrated range is 0 to 150 ppbv, with a recommended operating range of 1 to 150 ppbv, and trend indication continues to 1000 ppbv so that the nature of a process upset can be seen. Accuracy is ±1 ppbv or ±10% of reading, whichever is greater. The response to a 25 ppb step change in either direction reaches 80% in under 10 minutes.
The analyser works with most ultra high purity gases, including helium, argon, neon, xenon, krypton, nitrogen, hydrogen and oxygen. Changing gas is a single menu selection with no manual adjustment. A built-in online zero check confirms that measurements remain stable.
The sample is supplied at 138 to 345 kPa (20 to 50 psig) and vented to atmosphere, at a flow below 0.5 L/min. Inlet gas temperature can be 0 to 100 °C, with 60 °C ±5 °C being best. Pneumatics need instrument air at 80 to 100 psi, connected by 1/4 inch tubing. The unit is built for a 19 inch rack, measuring 480 x 177 x 509 mm and weighing 15.9 kg, with IP40 protection. The operator interface is a 4 line by 20 character LCD and an 18 key keypad.
Outputs are one self powered 4 to 20 mA analogue output into a 100 to 500 ohm load, RS-485 and RS-232 serial ports, and three alarm contacts for system, concentration and data valid, rated up to 30 V AC or 60 V DC. Power is 100 to 132 VAC or 230 VAC ±10%, 47 to 63 Hz, 185 W. Ambient temperature is 10 to 30 °C, with humidity up to 90% without condensation, and installation in Class I, Division 2, groups A to D is permitted.
Technical specifications
| Measured component | H2O |
|---|---|
| Principle | Quartz crystal microbalance (QCM) |
| Range | 0–150 ppbv; trend indication to 1000 ppbv |
| Detection limit | 1 ppbv |
| Accuracy | ±1 ppbv or ±10% of reading, whichever is greater |
| Response time | 80% of a 25 ppb step change in under 10 minutes |
| Inlet pressure | 138–345 kPa (20–50 psig) |
| Sample flow | Below 0.5 L/min |
| Inlet gas temperature | 0–100 °C |
| Analogue output | 4–20 mA into 100–500 ohm |
| Digital interfaces | RS-485, RS-232 |
| Alarms | System, concentration and data valid |
| Display | LCD, 4 x 20 characters |
| Supply voltage | 100–132 VAC or 230 VAC ±10%, 47–63 Hz, 185 W |
| Ambient temperature | 10–30 °C |
| Enclosure | 19 inch rack, IP40 |
| Dimensions (W x H x D) | 480 x 177 x 509 mm |
| Weight | 15.9 kg |
| Hazardous area | Class I, Division 2, groups A–D |
Request a price and lead time
5920 UHP Moisture Analyser for Ultra High Purity Gases, 0 to 150 ppbv. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which gases can the 5920 analyse?
Helium, argon, neon, xenon, krypton, nitrogen, hydrogen and oxygen of ultra high purity. The gas type is changed from the menu.
What happens if moisture rises above 150 ppbv?
The analyser continues to show the trend up to 1000 ppbv, so the nature of a process upset is recorded.
What sample conditions are needed?
A pressure of 138–345 kPa (20–50 psig), a flow below 0.5 L/min and an inlet gas temperature of 0–100 °C, vented to atmosphere.