5800 Trace Moisture Analyser for High-Purity Gases, 0.02–100 ppmv, with Moisture Generator
The 5800 analyser measures trace moisture in high-purity and specialty gases from 0.02 ppmv, with trend indication up to 1000 ppmv. A built-in moisture generator traceable to NIST checks the sensor's performance directly on the sample gas.
- Measuring range
- 0.02–100 ppmv, trend indication up to 1000 ppmv
- Detection limit
- 0.02 ppmv
- Accuracy
- ±20 ppbv or 5% of reading, whichever is greater
- Sensitivity
- 5 ppbv or 0.5% of range, whichever is greater
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The instrument is used wherever accurate measurement of trace moisture is required, including high-purity gas production, the production, storage and transmission of olefins, semiconductor gases and quality control of industrial gases. The main sectors are microelectronics, industrial gases and petrochemicals.
The analyser is compatible with almost all non-corrosive gases, including inert gases (He, Ar, Ne, Xe, Kr), O2, H2, N2, NO, CO, hydrocarbons, air and specialty gases such as SF6. A special measurement cell is needed for CO2, and compatibility with other gases is agreed with the manufacturer. Switching from one gas to another is quick.
The instrument uses a non-equilibrium measurement method and reaches 63% of a steady reading in less than five minutes, both when drying and when humidifying. An internal dryer conditions the reference gas. The moisture generator adds a known amount of water to the conditioned gas, and the resulting reference wet gas is fed to the sensor, with an alarm contact alerting the operator if the check fails. The check runs on a schedule or manually.
The measuring range is 0.02–100 ppmv, the detection limit is 0.02 ppmv (20 ppbv) and the calibrated range is 0.1–100 ppmv. The display can be set to show ppmv, ppbv or dew point temperature, the last requiring a pressure input. Two programmable 4–20 mA outputs into a 1200 ohm load can be assigned to any part of the scale, with one of them switching ranges automatically. A bidirectional RS485 port and three independent contact closures rated 32 V DC and 1 A non-inductive load are provided.
Inlet pressure is 1 to 6.89 bar (15 to 100 psig), exhaust is atmospheric, and sample flow is 600 mL/min at standard conditions. Sample gas temperature is 0 to 100 °C, and operating conditions are 4 to 40 °C at up to 90% relative humidity, non-condensing, in a non-corrosive atmosphere. The power supply is 85 to 265 V at 47 to 63 Hz. The case measures 432 × 132 × 381 mm, the rack-mount version is 221 mm high, and the weight is 16.79 kg.
Technical specifications
| Measuring range | 0.02–100 ppmv, trend indication up to 1000 ppmv |
|---|---|
| Detection limit | 0.02 ppmv |
| Accuracy | ±20 ppbv or 5% of reading, whichever is greater |
| Sensitivity | 5 ppbv or 0.5% of range, whichever is greater |
| Response time | 63% of a step change in either direction in less than 5 minutes |
| Compatible gases | He, Ar, Ne, Xe, Kr, O2, H2, N2, NO, CO, hydrocarbons, air, SF6 |
| Inlet pressure | 1 to 6.89 bar (15 to 100 psig) |
| Exhaust pressure | atmospheric |
| Sample flow | 600 mL/min at standard conditions |
| Sample gas temperature | 0 to 100 °C |
| Analogue outputs | 2 × 4 to 20 mA, 1200 ohm load |
| Digital interface | RS485, bidirectional |
| Alarms | 3 contact closures, 32 V DC, 1 A |
| Operating conditions | 4 to 40 °C, up to 90% relative humidity, non-condensing |
| Power supply | 85 to 265 V, 47 to 63 Hz |
| Dimensions (W × H × D) | 432 × 132 × 381 mm; rack-mount version height 221 mm |
| Net weight | 16.79 kg |
Part numbers
Show all part numbers (1)
5800
Request a price and lead time
5800 Trace Moisture Analyser for High-Purity Gases, 0.02–100 ppmv, with Moisture Generator. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
What is the lowest moisture level the 5800 can measure?
The detection limit is 0.02 ppmv, equal to 20 ppbv. The calibrated range is 0.1 to 100 ppmv, and the trend is shown up to 1000 ppmv.
Can the analyser measure moisture in CO2?
CO2 needs a special measurement cell. Inert gases, O2, H2, N2, CO, hydrocarbons and SF6 are measured with the standard version.
How are readings checked without an external reference?
The built-in generator adds a known amount of moisture to the conditioned gas and feeds it to the sensor. The check runs on a schedule or manually, and a failed check triggers the alarm contact.
Similar equipment from other manufacturers