3050 Series Trace Moisture Analysers Using Quartz Crystal Microbalance, from 0.01 ppmv
The 3050 series measures trace moisture in natural gas, hydrocarbon and industrial gases using quartz crystal microbalance (QCM) technology. A built-in verification system confirms accuracy without removing the analyser from service.
- Technology
- quartz crystal microbalance (QCM)
- Component measured
- H2O
- Range
- 0.01 to 2500 ppmv depending on model
- Accuracy
- from ±0.01 ppmv or ±10% (3050-TE)
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The 3050 series is built on quartz crystal microbalance (QCM) technology. Moisture adsorbed on a sensitive coating on the quartz crystal changes the frequency at which the crystal oscillates. The analysers measure continuously, from moderate levels down to below 1 ppmv, and respond quickly to both rises and falls in moisture.
Because the crystal is stable, factory recalibration is not needed. An internal moisture source allows accuracy to be checked on demand without removing the analyser. The reference gas is produced continuously from the sample itself through a dryer. Operators can view data and alarms and change settings from the built-in display.
The series comprises several models. The 3050-TE serves the inlet and outlet of turboexpanders in cryogenic plants, covering 0.01 to 100 ppmv with accuracy of ±0.01 ppmv or ±10%, and it includes a zero module and moisture generator. The 3050-SLR targets very low concentrations from 0.1 to 100 ppmv. The 3050-DO monitors the regeneration of molecular sieve dryers over 0.02 to 100 ppmv. The 3050-AP handles gases containing hydrogen sulphide, from 0.1 to 2500 ppmv.
The 3050-AMS and 3050-AM cover 0.035 to 100 ppmv and 0.1 to 100 ppmv respectively, and both offer online zero-gas verification. The 3050-RM is a 19 inch rack or bench-top unit, and the 3050-OLV comes in IP65 or NEMA 4X enclosures with an internal moisture generator. Sample pressure at the inlet is 1.38 to 3.45 bar (20 to 50 psi), with a regulator allowing up to 200 bar (3000 psi), and the sample temperature is 0 to 100 °C. Power is 120/240 V AC, 50/60 Hz, 150 W.
Outputs include an isolated 4 to 20 mA signal, RS-232 and RS-485 serial ports with Modbus RTU, Ethernet with Modbus TCP and a web interface, and USB. The analysers suit natural gas, inert gases, hydrogen, oxygen, nitrogen, air and sulphur hexafluoride. They are certified to ATEX and IECEx Ex db eb IIB+H2 T* Gb, and to UL/CSA Class I, Division 1, Groups B, C, D, T6 and Division 2, Groups A, B, C, D, T4.
Technical specifications
| Technology | quartz crystal microbalance (QCM) |
|---|---|
| Component measured | H2O |
| Range | 0.01 to 2500 ppmv depending on model |
| Accuracy | from ±0.01 ppmv or ±10% (3050-TE) |
| Inlet pressure | 1.38 to 3.45 bar (20 to 50 psi), up to 200 bar (3000 psi) with reducer |
| Exhaust pressure | 0 to 1 bar (0 to 15 psi) |
| Sample temperature | 0 to 100 °C |
| Sample flow | about 150 sccm, bypass about 1.0 L/min |
| Power | 120/240 V, 50/60 Hz, 150 W; optional 24 V DC |
| Outputs | isolated 4 to 20 mA, RS-232, RS-485 (Modbus RTU), Ethernet (Modbus TCP), USB |
| Alarm contacts | 2 to 4 |
| Enclosure | IP65 (Zone 1/Div 1), NEMA 4X (Div 2), 19 inch rack (3050-RM) |
| Explosion protection | ATEX/IECEx Ex db eb IIB+H2 T* Gb; UL/CSA Class I, Div 1, Groups B, C, D T6 |
| Verification | built-in moisture source, no removal needed |
Part numbers
Show all part numbers (8)
3050-TE, 3050-SLR, 3050-RM, 3050-OLV, 3050-DO, 3050-AP, 3050-AMS, 3050-AM
Request a price and lead time
3050 Series Trace Moisture Analysers Using Quartz Crystal Microbalance, from 0.01 ppmv. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which model measures the lowest moisture concentrations?
The 3050-TE, with a range of 0.01 to 100 ppmv and accuracy of ±0.01 ppmv or ±10%. It is designed for turboexpanders in cryogenic plants.
Is factory recalibration required?
No. The stability of the quartz crystal removes the need for factory recalibration, and accuracy is checked on site with the built-in moisture source.
Can the analyser measure gases containing sour components?
Yes. The series covers streams with up to 30% hydrogen sulphide or up to 100% carbon dioxide, and the 3050-AP is the model for hydrogen sulphide applications.
Similar equipment from other manufacturers