3050 Series Trace Moisture Analysers Using Quartz Crystal Microbalance, from 0.01 ppmv

The 3050 series measures trace moisture in natural gas, hydrocarbon and industrial gases using quartz crystal microbalance (QCM) technology. A built-in verification system confirms accuracy without removing the analyser from service.

Technology
quartz crystal microbalance (QCM)
Component measured
H2O
Range
0.01 to 2500 ppmv depending on model
Accuracy
from ±0.01 ppmv or ±10% (3050-TE)

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The 3050 series is built on quartz crystal microbalance (QCM) technology. Moisture adsorbed on a sensitive coating on the quartz crystal changes the frequency at which the crystal oscillates. The analysers measure continuously, from moderate levels down to below 1 ppmv, and respond quickly to both rises and falls in moisture.

Because the crystal is stable, factory recalibration is not needed. An internal moisture source allows accuracy to be checked on demand without removing the analyser. The reference gas is produced continuously from the sample itself through a dryer. Operators can view data and alarms and change settings from the built-in display.

The series comprises several models. The 3050-TE serves the inlet and outlet of turboexpanders in cryogenic plants, covering 0.01 to 100 ppmv with accuracy of ±0.01 ppmv or ±10%, and it includes a zero module and moisture generator. The 3050-SLR targets very low concentrations from 0.1 to 100 ppmv. The 3050-DO monitors the regeneration of molecular sieve dryers over 0.02 to 100 ppmv. The 3050-AP handles gases containing hydrogen sulphide, from 0.1 to 2500 ppmv.

The 3050-AMS and 3050-AM cover 0.035 to 100 ppmv and 0.1 to 100 ppmv respectively, and both offer online zero-gas verification. The 3050-RM is a 19 inch rack or bench-top unit, and the 3050-OLV comes in IP65 or NEMA 4X enclosures with an internal moisture generator. Sample pressure at the inlet is 1.38 to 3.45 bar (20 to 50 psi), with a regulator allowing up to 200 bar (3000 psi), and the sample temperature is 0 to 100 °C. Power is 120/240 V AC, 50/60 Hz, 150 W.

Outputs include an isolated 4 to 20 mA signal, RS-232 and RS-485 serial ports with Modbus RTU, Ethernet with Modbus TCP and a web interface, and USB. The analysers suit natural gas, inert gases, hydrogen, oxygen, nitrogen, air and sulphur hexafluoride. They are certified to ATEX and IECEx Ex db eb IIB+H2 T* Gb, and to UL/CSA Class I, Division 1, Groups B, C, D, T6 and Division 2, Groups A, B, C, D, T4.

Technical specifications

Technologyquartz crystal microbalance (QCM)
Component measuredH2O
Range0.01 to 2500 ppmv depending on model
Accuracyfrom ±0.01 ppmv or ±10% (3050-TE)
Inlet pressure1.38 to 3.45 bar (20 to 50 psi), up to 200 bar (3000 psi) with reducer
Exhaust pressure0 to 1 bar (0 to 15 psi)
Sample temperature0 to 100 °C
Sample flowabout 150 sccm, bypass about 1.0 L/min
Power120/240 V, 50/60 Hz, 150 W; optional 24 V DC
Outputsisolated 4 to 20 mA, RS-232, RS-485 (Modbus RTU), Ethernet (Modbus TCP), USB
Alarm contacts2 to 4
EnclosureIP65 (Zone 1/Div 1), NEMA 4X (Div 2), 19 inch rack (3050-RM)
Explosion protectionATEX/IECEx Ex db eb IIB+H2 T* Gb; UL/CSA Class I, Div 1, Groups B, C, D T6
Verificationbuilt-in moisture source, no removal needed

Part numbers

Show all part numbers (8)

3050-TE, 3050-SLR, 3050-RM, 3050-OLV, 3050-DO, 3050-AP, 3050-AMS, 3050-AM

Request a price and lead time

3050 Series Trace Moisture Analysers Using Quartz Crystal Microbalance, from 0.01 ppmv. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Which model measures the lowest moisture concentrations?

The 3050-TE, with a range of 0.01 to 100 ppmv and accuracy of ±0.01 ppmv or ±10%. It is designed for turboexpanders in cryogenic plants.

Is factory recalibration required?

No. The stability of the quartz crystal removes the need for factory recalibration, and accuracy is checked on site with the built-in moisture source.

Can the analyser measure gases containing sour components?

Yes. The series covers streams with up to 30% hydrogen sulphide or up to 100% carbon dioxide, and the 3050-AP is the model for hydrogen sulphide applications.