Explosion-Proof Oxygen Analyser 3020P, Calibration Range 0-25%, Three Programmable Spans
The 3020P measures the percentage of oxygen in process gases located in hazardous areas. It offers three programmable ranges starting at 0-1%, a calibration span of 0-25% and an enclosure approved for Class I, Division 1, Groups B, C and D.
- Measured component
- O2, percentage content
- Ranges
- 3 programmable, minimum 0-1%
- Calibration range
- 0-25%
- Accuracy
- ±2% FS at constant temperature; ±5% FS across the operating range
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
Operators fit this analyser to watch oxygen levels in inert gas blanketing of vessels, in air separation and liquefaction plants, during chemical reaction monitoring, in semiconductor manufacturing and in petrochemical process control. Quality assurance and gas analysis certification also make use of it.
The sensing element is a micro-fuel cell of class B-1 as standard, while A3, A5, B-3 and C-3 cells can be specified on order. A cell is expected to last eight months. An 8-bit CMOS microprocessor with 32 kB of RAM and 128 kB of ROM handles the signal, and the reading appears on a digital display. Accuracy is ±2% of full scale at constant temperature and ±5% of full scale across the operating temperature range. Sensitivity is 0.5% of full scale, and 90% of full scale is reached faster than 10 seconds at 25 °C.
Signals leave the unit as 0-1 V DC for concentration and 0-1 V DC for range identification by default. An isolated 4-20 mA DC output, with its own 4-20 mA range identifier, is an option. A bi-directional RS-232C link runs at 2400 baud. Alarms comprise one system failure contact and two fully programmable concentration set points, each with Form C contacts rated at 3 A.
Several variants suit different plants. The 3020P-C adds integral zero and calibration valves with separate ports for zero and span gas. The 3020P-F and 3020P-G are fitted with flame arrestors rated for Class I, Division 1, Groups C and D, and the -G also includes calibration valves. The 3020P-H and 3020P-I use flame arrestors for hydrogen service in Group B. The 3020P-S has a stainless steel sampling system and cell block, and the 3020P-M adds the isolated 4-20 mA output.
The enclosure is listed by UL and CSA and rated NEMA 4/7, with IP 66 ingress protection. The ATEX version is marked II 2G Ex db IIB+H2 T6 Gb and meets IEC 60079-0:2017 and IEC 60079-1:2014. Ambient operation runs from 0-50 °C on 115 or 230 V AC at 50 or 60 Hz, drawing up to 292 W. Sample flow must not exceed 1100 SCCM, the sensor inlet is limited to 0.86 psig (6 kPa), and applied gas pressure can reach 25 psig (172 kPa). The unit is not designed for gases with more than 21% oxygen. Mounting is by bulkhead, with 1/4-inch gas fittings and 3/4-inch or 1-inch NPT female ports to ANSI/ASME B1.20.1. The enclosure measures 25.63 in by 15.25 in by 11 in.
Technical specifications
| Measured component | O2, percentage content |
|---|---|
| Ranges | 3 programmable, minimum 0-1% |
| Calibration range | 0-25% |
| Accuracy | ±2% FS at constant temperature; ±5% FS across the operating range |
| Sensitivity | 0.5% FS |
| Response time | 90% FS in under 10 s at 25 °C |
| Sensor | micro-fuel cell class B-1; options A3, A5, B-3, C-3 |
| Analogue outputs | 0-1 V DC concentration plus 0-1 V DC range ID; optional isolated 4-20 mA DC |
| Interface | RS-232C, 2400 baud |
| Alarms | system failure; two set points with Form C contacts rated 3 A |
| Power supply | 115 / 230 V AC, 50-60 Hz, up to 292 W |
| Ambient temperature | 0-50 °C |
| Explosion protection | Class I, Division 1, Groups B, C, D; NEMA 4/7; II 2G Ex db IIB+H2 T6 Gb |
| Ingress protection (ATEX) | IP 66 |
| Gas connections | 1/4-inch fittings |
| Wetted parts | stainless steel 316, nylon |
| Dimensions (H × W × D) | 25.63 in × 15.25 in × 11 in |
Part numbers
Show all part numbers (12)
3020P, 3020P-C, 3020P-F, 3020P-G, 3020P-H, 3020P-I, 3020P-M, 3020P-S, 3020P (ATEX), 3020P (ATEX)-C, 3020P (ATEX)-S, 3020P (ATEX)-M
Request a price and lead time
Explosion-Proof Oxygen Analyser 3020P, Calibration Range 0-25%, Three Programmable Spans. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Can the 3020P be used on hydrogen-bearing gases?
Yes. The 3020P-H and 3020P-I carry flame arrestors for Group B (hydrogen) service, and the ATEX version is marked Ex db IIB+H2 T6 Gb.
Which signals can be wired into a control system?
As standard, 0-1 V DC for concentration and for range identification. The 3020P-M adds an isolated 4-20 mA output with range identification, and an RS-232C interface at 2400 baud is also available.
How long does the oxygen sensor last?
The micro-fuel cell of class B-1 is expected to last eight months. Cells A3, A5, B-3 and C-3 can be ordered instead.
Similar equipment from other manufacturers