The 2850 analyser monitors moisture in clean gases after adsorption dryers, in instrument air systems and at air separation plants. Its operating range is 0.1 to 1000 ppmv, with a factory calibration of 1 to 100 ppmv, and it reaches 63% of a step change in less than five minutes.

Operating range
0.1–1000 ppmv
Factory calibrated range
1–100 ppmv
Accuracy
±0.05 ppmv or ±5% of reading, whichever is greater
Sensitivity
20 ppbv or 0.5% of range, whichever is greater

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The main purpose of the instrument is monitoring of process adsorption dryers, instrument air dryers and air separation units, as well as quality control in industrial gas production and cryogenic air separation. The manufacturer lists petrochemicals, chemicals, refining, LCD and OLED display manufacturing, semiconductor production, hydrogen-cooled generators and heat treatment of steels and other metals among the application areas.

The analyser uses a non-equilibrium measurement method, so it responds equally quickly to drying and to humidifying of the gas. The reference gas is dried to no more than 10 ppbv. Between 100 ppbv and 1 ppmv the reading serves as a sensitive trend indicator, while values below 0.1 ppmv are not shown on the display.

A built-in moisture generator traceable to NIST adds a known amount of water to the conditioned sample gas and feeds it to the sensor. If the check is not passed, an alarm contact alerts the operator. The check runs on a programmable schedule or manually.

The four-line, 20-character vacuum fluorescent display shows moisture in ppmv or as a dew point, together with timer and instrument status. Two programmable 4–20 mA outputs into a 1200 ohm load can be assigned to any part of the calibrated range. A bidirectional RS-485 port is provided, along with three pairs of contacts for system alarm, range or calibration alert and concentration alert, rated 32 V and 1 A non-inductive, failing safe by default.

Inlet pressure is 1 to 6.89 bar (15 to 100 psig), operating and exhaust pressures are atmospheric, and sample flow is below 600 mL/min at standard conditions. Sample gas temperature is 0 to 100 °C and ambient conditions are 4 to 40 °C at up to 90% relative humidity, non-condensing. Installation is permitted at altitudes up to 2000 m. The power supply is 85 to 265 V at 47 to 63 Hz. The unit is available for bench mounting or in a 19-inch rack, with dimensions of 432 × 132 × 381 mm, a rack-mount height of 221 mm and a weight of 17 kg. It is rated NEC Division 2 with CE conformity.

Technical specifications

Operating range0.1–1000 ppmv
Factory calibrated range1–100 ppmv
Accuracy±0.05 ppmv or ±5% of reading, whichever is greater
Sensitivity20 ppbv or 0.5% of range, whichever is greater
Response time63% of a step change in less than 5 minutes
Inlet pressure1–6.89 bar (15–100 psig)
Operating and exhaust pressureatmospheric
Sample flowbelow 600 mL/min at standard conditions
Sample gas temperature0 to 100 °C
Operating conditions4 to 40 °C, up to 90% relative humidity, altitudes up to 2000 m
Analogue outputs2 × 4 to 20 mA, 1200 ohm load
Digital interfaceRS-485, bidirectional
Alarms3 pairs of contacts, 32 V, 1 A
Power supply85 to 265 V, 47 to 63 Hz
Dimensions (W × H × D)432 × 132 × 381 mm; rack-mount version height 221 mm
Net weight17 kg
Electrical classificationNEC Division 2, CE

Part numbers

Show all part numbers (1)

2850

Request a price and lead time

2850 Clean Gas Moisture Analyser for Dryers and Air Separation, 0.1–1000 ppmv. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

In which range does the 2850 give its specified accuracy?

The accuracy of ±5% of reading or ±0.05 ppmv applies within the factory calibrated range of 1 to 100 ppmv. From 100 ppbv to 1 ppmv, and above 100 ppmv, the instrument shows a trend.

How quickly does the instrument respond to a change in moisture?

It reaches 63% of a step change in less than five minutes. The response is the same when drying and when humidifying the gas.

Can the analyser be built into a cabinet?

A 19-inch rack version with a height of 221 mm is available. The bench version measures 432 × 132 × 381 mm.

Similar equipment from other manufacturers

LET’S CONNECT

Let’s start with your challenge.

Complete a short brief. We will prepare an email you can send from your own email app.

Call us044 339-50-2009:00–18:00
USK ENERGY / SEARCH

Search

Solutions, specialist directions and guides