Model 2020 Thermal Conductivity Gas Analyser in Explosion-Proof Enclosure, Ranges to 0 to 100%
Model 2020 is a thermal conductivity analyser for continuous measurement of a component in a gas mixture in hazardous areas. The explosion-proof enclosure carries the marking II 2 G Ex db IIB+H2 T6..T5 Gb, with ranges from 0 to 9999 ppm to 0 to 100%.
- Measuring principle
- thermal conductivity
- Ranges
- 0 to 9999 ppm to 0 to 100%
- Sensor temperature control
- 60 °C, to within 0.1 °C
- Analogue outputs
- 0 to 1 V DC, isolated 4 to 20 mA
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The analyser determines the concentration of a component in a gas mixture from changes in its thermal conductivity. It is installed on process plants where the sample may contain toxic or flammable gases, which is why the basic 2020 is built in an explosion-proof enclosure. It is approved for Class 1, Division 1, groups B, C and D, and holds ATEX certificate ITS19ATEX104845X and IECEx certificate ITS 19.0013X.
The thermal conductivity cell is regulated at 60 °C to within 0.1 °C, which reduces the effect of ambient temperature changes. Signal processing is handled by an 8031 microprocessor with 32 kB of RAM and 128 kB of ROM. Readings are shown on a two-line alphanumeric display, and the ranges are configurable from 0 to 9999 ppm to 0 to 100%.
The analyser has a 0 to 1 V DC analogue output and an isolated 4 to 20 mA output. Two concentration alarm relays and one system fault relay have changeover contacts (Form C).
The sample is supplied at a flow of no more than 1100 SCCM, with sensor inlet pressure up to 0.86 psig (6 kPa) and a maximum gas pressure on the instrument of 25 psig (172 kPa). The external breather and drain valve and the gas restrictors form part of the explosion protection and must not be altered. The analyser is not permitted for gases enriched with oxygen above 21%.
Options are denoted by suffixes: L for a flow regulation panel, C for calibration valves, R for a sealed reference cell, FAB for flame arrestors for group B, and FACD for groups C and D. Electrical connections are made through 3/4 or 1 inch NPT threaded openings to ANSI/ASME B1.20.1 with Ex d IIC glands, with at least five full threads engaged. The enclosure is earthed with an M6 bolt and 10 AWG or larger wire.
Power is 115 or 230 V AC, 50 or 60 Hz, consuming up to 220 W. Operating conditions are 0 to 50 °C with IP 66 protection. For general industrial areas the model 2000B is available in a NEMA-4 enclosure, and for a remote layout the model 2010 has an explosion-proof analytical unit with a separate control unit.
Technical specifications
| Measuring principle | thermal conductivity |
|---|---|
| Ranges | 0 to 9999 ppm to 0 to 100% |
| Sensor temperature control | 60 °C, to within 0.1 °C |
| Analogue outputs | 0 to 1 V DC, isolated 4 to 20 mA |
| Alarm relays | two concentration relays, one system fault relay (Form C) |
| Display | two-line alphanumeric |
| Microprocessor | 8031, RAM 32 kB, ROM 128 kB |
| Power supply | 115 or 230 V AC, 50/60 Hz, up to 220 W |
| Sample flow | up to 1100 SCCM |
| Sensor inlet pressure | up to 0.86 psig (6 kPa) |
| Maximum gas pressure | 25 psig (172 kPa) |
| ATEX explosion protection | II 2 G Ex db IIB+H2 T6..T5 Gb |
| IECEx explosion protection | Ex db IIB+H2 T6..T5 Gb |
| North American version | Class 1, Div 1, groups B, C, D |
| Operating conditions | 0 to 50 °C, IP 66 |
| Restriction | not for gases with oxygen above 21% |
Part numbers
Show all part numbers (11)
2020-L, 2020-C, 2020-R, 2020-FAB, 2020-L-FAB, 2020-C-FAB, 2020-R-FAB, 2020-FACD, 2020-L-FACD, 2020-C-FACD, 2020-R-FACD
Request a price and lead time
Model 2020 Thermal Conductivity Gas Analyser in Explosion-Proof Enclosure, Ranges to 0 to 100%. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Can the 2020 be used with oxygen-enriched gases?
No, the analyser is not permitted for gases containing more than 21% oxygen. The explosion protection is rated for gases up to group IIB+H2.
Which outputs does the analyser provide for a control system?
A 0 to 1 V analogue output, an isolated 4 to 20 mA output, two concentration alarm relays and a system fault relay with Form C contacts.
Is there a version for non-hazardous areas?
Yes, the model 2000B in a NEMA-4 enclosure is for general industrial areas, while the model 2010 has a remote layout with a separate control unit.