2000A Thermal Conductivity Gas Analyser, 3 Ranges, 4-20 mA and RS-232 Outputs
The 2000 series is a panel-mounted microprocessor analyser that measures the concentration of a component in a binary or multi-component gas mixture by comparing the thermal conductivity of the sample with a reference gas. It offers three user ranges, automatic range selection, 4-20 mA and 0-1 V analogue outputs, and an RS-232 port.
- Measuring principle
- thermal conductivity, Wheatstone bridge
- Sensor
- standard TC cell, four-filament detector
- Ranges
- 3 user ranges plus calibration range, auto-ranging
- Accuracy
- ±1% FS at constant temperature; ±5% FS over operating temperature range
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
Each cell has two chambers, one holding a reference gas of known composition, sealed or flowing, and the other the sample. Each chamber contains a pair of heated filaments forming part of a Wheatstone bridge. A change in sample composition alters the heat drawn from the filaments, unbalancing the bridge, so the imbalance signal is proportional to the change in composition. The standard cell uses a four-filament detector with a nickel-plated brass block, nickel alloy filaments and stainless steel tubing and end plates.
The analyser can measure one gas in a mixture of two, the concentration of a gas in a set background mixture, or the purity of a stream with impurities. Linearisation for many gases and mixtures is stored in memory, and non-standard tasks can be added by the manufacturer or by the user. Typical uses include gas purity control in air separation, calibration and breathing gas blends, hydrogen monitoring in refinery recycle streams, reforming, synthesis gas and protective atmospheres in annealing furnaces.
Three independent ranges can be set for three gases or three scales of one gas, with a separate recalibration range. The display is a 2-line backlit LCD with a 5-digit LED readout. Outputs are two 0-1 V DC and two isolated 4-20 mA DC signals for concentration and range identification, with loads up to 1000 Ohm, plus a full-duplex RS-232 port. Two programmable concentration set points have Form C 3 A contacts, and a separate contact signals system failure.
Accuracy is ±1% of full scale for most binary mixtures at constant temperature, and ±5% of full scale across the 0 to 50 °C operating range once temperature has settled. Zero and span drift is ±1% of full scale, and the response time is 90% in under 10 seconds at 100 sccm. The sample flow is 0.1 to 0.4 SCFH at 5 to 50 psig. Connections are 1/4-inch compression fittings, and wetted parts are stainless steel, nickel and gold.
The 2000AEU version is for panel and rack mounting in indoor areas without explosion risk. It meets EU EMC requirements, is classed as ISM Class A Group I and complies with EN55011 and EN50082-2. Optional FM approval covers Class I, Div 2, and no purge version exists. Power is 110 V AC, 50 to 60 Hz, with 220 V AC optional. For hazardous areas the 2000B is available, as are wall-mounted 2010A, 2010B and 2020 models.
Technical specifications
| Measuring principle | thermal conductivity, Wheatstone bridge |
|---|---|
| Sensor | standard TC cell, four-filament detector |
| Ranges | 3 user ranges plus calibration range, auto-ranging |
| Accuracy | ±1% FS at constant temperature; ±5% FS over operating temperature range |
| Zero and span drift | ±1% FS |
| Response time | 90% in under 10 s at 100 sccm |
| Operating temperature | 0–50 °C |
| Analogue outputs | 2 × 0–1 V DC; 2 × 4–20 mA DC isolated |
| Load impedance | up to 1000 Ohm |
| Interface | RS-232, full duplex |
| Alarms | 2 set points, Form C 3 A contacts; system failure contact |
| Display | 2-line backlit LCD; 5-digit LED |
| Cell material | nickel-plated brass, nickel alloy filaments, stainless steel |
| Sample flow | 0.1–0.4 SCFH |
| Sample pressure | 5–50 psig |
| Connections | 1/4-inch compression fittings, 6 mm adapters optional |
| Power | 110 V AC, 50–60 Hz (220 V AC optional) |
| Area classification | general purpose, FM Class I, Div 2 optional |
| Standards | EN55011 Class A Group I, EN50082-2 |
Part numbers
Show all part numbers (2)
2000A, 2000AEU
Request a price and lead time
2000A Thermal Conductivity Gas Analyser, 3 Ranges, 4-20 mA and RS-232 Outputs. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which tasks can the 2000A handle?
It measures a component in a binary mixture, a gas in a given background mixture, or the purity of a stream with impurities. Linearisation is built in for many gases and mixtures, and non-standard tasks can be added by the manufacturer or by the user.
Can the 2000A be used in a hazardous area?
The base version is for general purpose areas, with an optional FM approval for Class I, Div 2. There is no purge version. For Class I, Div 1 the series includes the 2000B and 2010B.
Which signals does the analyser send to the control system?
Two 0–1 V DC outputs and two isolated 4–20 mA DC outputs for concentration and range identification, an RS-232 port and Form C 3 A contacts.
Similar equipment from other manufacturers