International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

NOMAD-Direct portable 2-channel fibre optic thermometer for probe checks, −20…150 °C

The NOMAD-Direct is a portable two-channel instrument for checking that fibre optic temperature probes in transformer windings are working and correctly installed during manufacture, assembly and testing. Its accuracy is ±1 °C, resolution is 0.1 °C and response time is 0.5 s.

Type
portable fibre optic temperature tester
Channels
2
Calibrated range
−20…150 °C
Error
±1 °C

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The instrument runs from a built-in lithium-ion battery for up to 3 hours. It charges through a USB Type-C port at 5 V DC, up to 1.1 A, and if the battery is completely flat it should be charged for at least 15 minutes before switching on. It measures 202 x 170 x 50 mm and weighs 1 kg.

Designed for indoor use at 0 to 45 °C with humidity up to 95 % non-condensing, the unit has a pollution degree of PD2 and can be stored at −20 to 50 °C. It conforms to CSA 61010-1 and UL 61010-1.

The instrument is used at the transformer factory and on site: after the probes are placed in the winding, at key assembly stages, before commissioning and during maintenance. Repeated quick readings confirm that a probe is undamaged and correctly fitted before the transformer is energised. The check is recommended to be repeated more than ten times per unit, which reduces rework. NOMAD-Direct replaces the earlier Nomad-Touch (NMT) model.

A bright display shows temperature in real time, along with probe health indicators such as signal level, CCD time, lamp and ratio, which give a quick assessment of probe suitability. The instrument is calibrated at the factory, so no separate system calibration is needed. The calibrated range is −20 to 150 °C, with ±1 °C error, 0.1 °C resolution, 0.5 s response time and 2 channels.

Dielectric fibre optic probes carry no current, so readings are insensitive to electromagnetic and radio frequency interference and stay stable near high-voltage windings, which allows work at substations and power stations with strong fields. The unit works with all of the maker's T1 and T2 fibre optic probes. It is a test instrument and is not designed for continuous or primary monitoring of transformer temperature.

Technical specifications

Typeportable fibre optic temperature tester
Channels2
Calibrated range−20…150 °C
Error±1 °C
Resolution0.1 °C
Response time0.5 s
Compatible probesT1, T2
Operating temperature0…45 °C
Humidityup to 95 % non-condensing
Storage temperature−20…50 °C
Pollution degreePD2, indoor use only
Batterylithium-ion, up to 3 h
ChargingUSB Type-C, 5 V DC, up to 1.1 A
Dimensions202 x 170 x 50 mm
Weight1 kg
StandardsCSA 61010-1, UL 61010-1

Part numbers

Show all part numbers (2)

NOMAD-DIRECT, NOMAD-Direct-CN

Request a price and lead time

NOMAD-Direct portable 2-channel fibre optic thermometer for probe checks, −20…150 °C. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Can NOMAD-Direct be used for continuous monitoring of a transformer?

No, it is a test instrument for checking probes during manufacture, installation and maintenance, not for continuous monitoring.

Which probes can be checked with this instrument?

All of the maker's T1 and T2 fibre optic probes, with up to 2 channels at the same time.

How long does the instrument run on its battery?

Up to 3 hours. If the battery is completely flat, it is charged through USB Type-C for at least 15 minutes before being switched on.