PD-Guard online monitor for partial discharge, 4 or 8 channels, 50 kHz to 1000 MHz

PD-Guard is a fixed analyser for continuous monitoring of partial discharge in transformers, rotating machines, cables and switchgear, including gas-insulated and metal-clad units. It analyses PD and electromagnetic interference in the HF, VHF and UHF ranges and issues warnings of insulation deterioration.

Inputs
4 or 8, multiplexed
Connector
BNC
Impedance
50 ohms
Maximum input
+10 dBm (+25 dBm with optional attenuator)

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The PD-Guard is installed on critical assets for continuous monitoring of partial discharge (PD). Its applications are power transformers, rotating machines, cables and switchgear, including gas-insulated and metal-clad units. Local alarms are provided through relay outputs, and data and notifications are sent over standard interfaces to networked supervisory systems. It is configured from a computer, tablet or any device with a web browser.

The analysis covers Qiec measurements, consistent with IEC 60270, and Qhf for field conditions, along with phase-resolved analysis and a PAPR indicator that tracks the trend in PD and interference intensity and shows when detailed diagnosis is needed. Detectors are peak, quasi-peak and average, and the quasi-peak detector conforms to CISPR 16-1-1:2010. A built-in expert system learns PD behaviour and reports changes in frequency and severity.

Inputs are four or eight multiplexed channels with BNC connectors, an input impedance of 50 Ω and a dynamic range of 60 dB. The maximum input is +10 dBm for reading, or +25 dBm with the optional attenuator. The first RFI tuner covers 50 kHz to 50 MHz, the second covers 50 MHz to 1000 MHz with ±100 kHz accuracy, and the EMI mode runs from 50 kHz to 100 MHz.

The monitor works with a range of sensors: antennas for airborne PD, clip-on high-frequency current transformers for single or bundled conductors, and UHF drain-valve probes for in-tank use. Clip-on HFCT sensors cover 10 kHz to 200 MHz and suit lines carrying up to 300 A. UHF drain-valve sensors fit DN 50 or DN 80 valves on transformers, plate sensors suit transformers without oil flanges, and spacer sensors suit GIS. A pulse generator verifies sensitivity in the UHF and VHF range.

The host runs an ARMv8 processor at 1.2 GHz with 1 GB RAM, 4 GB flash and 32 GB eMMC for applications and data. Interfaces include an isolated RS485 port for Modbus or DNP3, and 10/100 BaseT Ethernet for DNP3, Modbus and HTTP. IEC 61850 is an option. The status relay is rated 240 V AC at 5 A. The anodised aluminium housing measures 200 x 330 x 82 mm and weighs 2 kg. It runs from 24 V DC @ 1 A, with an operating range of -20 to +50 °C, and it works stand-alone, networked into a SCADA system, or as part of a doblePRIME platform.

Technical specifications

Inputs4 or 8, multiplexed
ConnectorBNC
Impedance50 ohms
Maximum input+10 dBm (+25 dBm with optional attenuator)
Dynamic range60 dB
Detectorspeak, quasi-peak, average
Sweep processingcontinuous, average, max hold, differential
Tuner 1 band50 kHz to 50 MHz, RBW 9 kHz / 120 kHz
Tuner 2 band50 MHz to 1000 MHz, ±100 kHz, RBW 120 kHz / 1 MHz / 6 MHz
EMI mode50 kHz to 100 MHz, RBW 9 kHz / 120 kHz
ProcessorARMv8 1.2 GHz
Memory1 GB RAM, 4 GB flash, 32 GB eMMC
Interfacesisolated RS485 (Modbus, DNP3), Ethernet 10/100 BaseT (DNP3, Modbus, HTTP), IEC 61850 option
Status relay240 V AC, 5 A
StandardsIEC 60270; CISPR 16-1-1, edition 2010
Power supply24 V DC @ 1 A
Operating temperature-20 °C to +50 °C
Storage temperature-20 °C to +70 °C
Dimensions (H x W x D)200 x 330 x 82 mm
Weight2 kg

Request a price and lead time

PD-Guard online monitor for partial discharge, 4 or 8 channels, 50 kHz to 1000 MHz. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Which equipment can be monitored?

Power transformers, rotating machines, cables and switchgear, including gas-insulated and metal-clad units.

Which sensors can be connected?

Antennas for airborne PD, HFCT current transformers, UHF drain-valve sensors for DN 50 or DN 80 valves, plate sensors, and sensors for GIS and bushing tap connectors.

Do the measurements conform to IEC 60270?

Yes. The Qiec measurement is consistent with IEC 60270, and the quasi-peak detector conforms to CISPR 16-1-1:2010.

Similar equipment from other manufacturers

LET’S CONNECT

Let’s start with your challenge.

Complete a short brief. We will prepare an email you can send from your own email app.

Call us044 339-50-2009:00–18:00
USK ENERGY / SEARCH

Search

Solutions, specialist directions and guides