TIY programmable isolated thermocouple transmitter, 4-20mA, 12-42DC, push button calibration
The TIY converts a thermocouple or millivolt input into 4-20mA, linear with temperature. Quick Ranging push buttons, an integral indicator, 2-wire 12-42DC power and 500Vrms isolation.
- Type
- Programmable isolated thermocouple transmitter
- Input
- ISA thermocouples (J, K, E, T, R, S, N) and others, mV
- Temperature range
- -190°C to +1820°C, no minimum span
- Output
- 4-20mA, linear with temperature or millivolt input
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The TIY is a digital/analogue hybrid: a microprocessor-based, 2-wire, programmable isolated transmitter. It accepts a non-linear temperature input from standard ISA thermocouples (types J, K, E, T, R, S and N) and other types, as well as from millivolt sources. The 4-20mA output is linear with temperature or with the millivolt input.
Setup is made in the field with tactile push buttons and quick-change solderless jumpers, so no hand-held configurator is needed. Thermocouple type, range, upscale or downscale drive and linearisation are selected there. Zero and span are set by Quick Ranging while reading the integral indicator, with no calibrator. The temperature range is -190°C to +1820°C with no minimum span, and automatic digital reference junction compensation maintains long-term accuracy.
The indicator reads the input in °C, °F or mV, shows zero and span adjustments, and keeps showing the process temperature even when the output goes overrange or underrange. Self-diagnostics sense sensor failure and errors in calibration, memory and unit function, and indicate them on the display.
Output accuracy is ±0.05% of span, ripple is below 10mVp-p up to 120Hz across a 250 ohm load and common mode rejection is at least 120dB at 50Hz/60Hz. 500Vrms galvanic isolation between input and output prevents false signals from ground loops. The unit withstands ±42V on the input and ±60V on the output without damage, and output current limiting is 125% of span typical and 130% of span maximum.
The housing is a hockey puck type with a DIN rail clip. For field mounting the unit can be ordered in an explosion-proof enclosure, and the DIN-style housing suits high-density cabinet installation. The rating is Non-Incendive/Div. 2, Zone 2. The unit is not for use in SIS and is RoHS compliant.
Technical specifications
| Type | Programmable isolated thermocouple transmitter |
|---|---|
| Input | ISA thermocouples (J, K, E, T, R, S, N) and others, mV |
| Temperature range | -190°C to +1820°C, no minimum span |
| Output | 4-20mA, linear with temperature or millivolt input |
| Power | 12-42DC, 2-wire, loop powered |
| Output accuracy | ±0.05% of span |
| Isolation | 500Vrms galvanic isolation between input and output |
| Common mode rejection | At least 120dB at 50Hz/60Hz |
| Ripple | Less than 10mVp-p (up to 120Hz) across a 250 ohm load |
| Output current limiting | 125% of span typical, 130% of span maximum |
| Operating limits | No damage up to ±42V on input and ±60V on output |
| Burnout protection | Upscale (standard), downscale by solderless jumper |
| Housing | Hockey puck housing with DIN rail clip |
| Integral display | Yes |
| Configuration | Push buttons, solderless jumpers |
| Hazardous area certifications | Non-Incendive/Div. 2, Zone 2 |
Part numbers
Show all part numbers (1)
TIY/PRG/4-20mA/12-42DC [DN]
How it compares with similar models
| Parameter | TIY programmable isolated thermocouple transmitter, 4-20mA, 12-42DC, push button calibration | RIY programmable isolated RTD transmitter, 4-20mA, 12-42DC, push button setup, display | TDZ3 dual input HART temperature transmitter, 4-20mA, 12-42DC loop powered | STZ functional safety dual input HART temperature transmitter, SIL 2/3, 4-20mA, 12-42DC |
|---|---|---|---|---|
| Type | Programmable isolated thermocouple transmitter | Programmable isolated RTD transmitter | Dual input smart HART temperature transmitter | Functional safety dual input smart HART temperature transmitter |
| Input | ISA thermocouples (J, K, E, T, R, S, N) and others, mV | 2-, 3- or 4-wire RTD or resistance source | Dual sensor, programmable: RTD, T/C, mV, ohms, potentiometer | One or two sensors: RTD (14 types), T/C (9 types), mV, ohms, potentiometer |
| Output | 4-20mA, linear with temperature or millivolt input | 4-20mA, linear with temperature or ohm input, direct or reverse | 4-20mA with HART | 4-20mA with HART (HART 7) |
| Isolation | 500Vrms galvanic isolation between input and output | Fully isolated | Fully isolated | Fully isolated |
| Integral display | Yes | Yes, 4-digit LCD | Yes | Yes |
| Configuration | Push buttons, solderless jumpers | Push buttons, switches | PC, USB cable, HART | PC, USB cable, HART |
Request a price and lead time
TIY programmable isolated thermocouple transmitter, 4-20mA, 12-42DC, push button calibration. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which inputs does the TIY accept?
ISA thermocouples (types J, K, E, T, R, S and N) and other types, plus millivolt sources, over -190°C to +1820°C with no minimum span.
How is the TIY calibrated?
Zero and span are set with the push buttons by Quick Ranging, reading the integral indicator, with no calibrator.
How much isolation is there between input and output?
500Vrms galvanic isolation, which prevents false signals due to ground loops.
Similar equipment from other manufacturers