T2X (TEMPTATION 2000) PC-programmable RTD temperature transmitter, 4-20mA, 8-42DC, non-isolated
The T2X is set up from a PC in a minute or less, accepts Pt100, Pt1000, Ni120 and Cu10 RTDs and 0-2200 ohm inputs, and gives a linear non-isolated 4-20mA. It is a 2-wire unit powered at 8-42DC.
- Type
- Non-isolated PC-programmable RTD temperature transmitter
- Input
- 2-, 3- and 4-wire RTD (Pt100, Pt1000, Ni120, Cu10), 0-2200 ohm, potentiometer
- Output
- 4-20mA, linear with temperature, non-isolated
- Power
- 8-42DC, 2-wire, loop powered
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The TEMPTATION 2000 (model T2X) is an affordable temperature transmitter for basic sensing applications. It is PC-programmable in a minute or less and handles 2-, 3- and 4-wire RTDs (Pt100, Pt1000, Ni120 and Cu10), 0-2200 ohm inputs and potentiometers. The 4-20mA output is linear with temperature and non-isolated, ready for direct interface with a monitoring or control system.
The transmitter is 2-wire and loop powered at 8-42DC. Load capability is 670 ohms at 24V, calculated as (supply voltage - 8V) / 0.024A. Setup is made in a single window of the PC software and stored in non-volatile memory. Total Sensor Diagnostics show which RTD wire has failed and drive the output upscale to 24mA or downscale to 3.3mA. Damping is selectable from 0-5 seconds.
Input accuracy is ±0.2°C for platinum, ±0.16°C for nickel, ±1.2°C for copper and ±0.4Ω for ohms, and output accuracy is ±0.05% of span. Stability is ±0.12% over 1 year, ±0.21% over 3 years and ±0.27% over 5 years. The output updates at least 8 times per second, with a response time of 256msec typical and 300msec maximum. The ambient operating range is -40°C to +85°C, with relative humidity of 0-95%, non-condensing.
The T2X is offered in a standard head-mount housing, in field-ready NEMA 4X, IP66 connection heads and with a DIN rail option, and complete sensor and thermowell assemblies are available. This version is the HPP with no enclosure, electronics only, measuring 52mm x 49mm x 26mm and weighing 101 g (3.6 oz). It is RFI/EMI resistant. For hazardous areas the rating is Non-Incendive/Div. 2, Zone 2. The unit is not for use in SIS and is RoHS compliant.
Technical specifications
| Type | Non-isolated PC-programmable RTD temperature transmitter |
|---|---|
| Input | 2-, 3- and 4-wire RTD (Pt100, Pt1000, Ni120, Cu10), 0-2200 ohm, potentiometer |
| Output | 4-20mA, linear with temperature, non-isolated |
| Power | 8-42DC, 2-wire, loop powered |
| Input accuracy | Platinum ±0.2°C, nickel ±0.16°C, copper ±1.2°C, ohms ±0.4Ω |
| Output accuracy | ±0.05% of span |
| Stability | 1 year ±0.12%, 3 years ±0.21%, 5 years ±0.27% |
| Measurement cycle | Output updates at least 8 times per second |
| Output response time | 256msec typical, 300msec maximum (10% to 90%) |
| Damping | 0-5 seconds, user selectable |
| Burnout protection | Upscale to 24mA or downscale to 3.3mA |
| Ambient operating range | -40°C to +85°C (-40°F to +185°F) |
| Dimensions | 52mm x 49mm x 26mm, mounting holes 7mm |
| Weight | HPP: 101 g (3.6 oz) |
| Configuration | PC, USB cable |
| Hazardous area certifications | Non-Incendive/Div. 2, Zone 2 |
Part numbers
Show all part numbers (1)
T2X/PRG/4-20mA/8-42DC [HPP]
How it compares with similar models
| Parameter | T2X (TEMPTATION 2000) PC-programmable RTD temperature transmitter, 4-20mA, 8-42DC, non-isolated | TIY programmable isolated thermocouple transmitter, 4-20mA, 12-42DC, push button calibration | SPT site-programmable isolated temperature transmitter, T/C, RTD, ohms, mV, 4-20mA or 1-5V | TDY PC-programmable temperature transmitter with large display, 4-20mA, 10-42DC |
|---|---|---|---|---|
| Type | Non-isolated PC-programmable RTD temperature transmitter | Programmable isolated thermocouple transmitter | Site-programmable isolated temperature transmitter | PC-programmable temperature transmitter with display |
| Input | 2-, 3- and 4-wire RTD (Pt100, Pt1000, Ni120, Cu10), 0-2200 ohm, potentiometer | ISA thermocouples (J, K, E, T, R, S, N) and others, mV | T/C (all common ISA), 2-, 3- or 4-wire RTD, ohms, mV | — |
| Output | 4-20mA, linear with temperature, non-isolated | 4-20mA, linear with temperature or millivolt input | 4-20mA (source or sink) or 1-5V, switch selectable | 4-20mA, linear, scalable, 4-20mA only |
| Power | 8-42DC, 2-wire, loop powered | 12-42DC, 2-wire, loop powered | 22-300VDC or 90-260VAC, auto-sensing | 10-42DC, 2-wire, loop powered |
| Output accuracy | ±0.05% of span | ±0.05% of span | ±0.03% of output span | — |
| Configuration | PC, USB cable | Push buttons, solderless jumpers | Push button, on-screen menu | PC, USB cable |
Request a price and lead time
T2X (TEMPTATION 2000) PC-programmable RTD temperature transmitter, 4-20mA, 8-42DC, non-isolated. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which sensors does the T2X accept?
2-, 3- and 4-wire RTDs (Pt100, Pt1000, Ni120, Cu10), 0-2200 ohm inputs and potentiometers.
What happens if an RTD wire fails?
Total Sensor Diagnostics show which wire has failed, and the output is driven upscale to 24mA or downscale to 3.3mA.
How stable is the T2X?
Stability is ±0.12% over 1 year, ±0.21% over 3 years and ±0.27% over 5 years.