TCM temperature concentrator module, 16 channels, HART, 15-30DC, intrinsically safe, DIN rail
The TCM has 16 channels, each configurable for RTD, T/C, millivolt or resistance/potentiometer signals, and sends the data over HART. Intrinsically safe and non-incendive, 15-30DC supply, DIN rail housing.
- Type
- Temperature concentrator module, 16 channels
- Inputs
- 16 channels, each configurable: RTD, T/C, mV, resistance/potentiometer
- RTD
- 2-, 3-, 4-wire; Pt, Cu, Ni
- Thermocouples
- J, K, E, T, R, S, N, B, C
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The TCM Temperature Concentrator Module has 16 channels that can be individually configured to accept an RTD, T/C, millivolt and/or resistance/potentiometer signal. It is approved for intrinsic safety and non-incendive use. All data is transmitted over HART to a HART host, so a dedicated transmitter and twisted wire pair, or expensive sensor extension wire, is not needed for each measurement.
The inputs accept any combination of RTD (2-, 3-, 4-wire; Pt, Cu, Ni), thermocouples J, K, E, T, R, S, N, B and C, resistance/potentiometer (0-4000 ohms) and millivolt (-50 to 1000mV) signals. In intrinsically safe installations only one isolated I.S. barrier is needed for all 16 points.
For MODBUS RTU the TCM is used with the HMC HART-to-MODBUS converter, and for MODBUS/TCP or HART-IP with the HES HART-to-Ethernet Gateway System. The HMC and the single channel HES can communicate with two TCMs only. The module can be installed in Class I, II, III, Division 1 (Zone 0/1) hazardous areas.
The supply is 15-30DC and the digital output is HART only. The metal housing is RFI/EMI resistant and snaps onto DIN-style rails. Configuration is from a PC over a USB cable. There is no integral display and no alarm or relay output, and the module is not for use in SIS. It is RoHS compliant.
Technical specifications
| Type | Temperature concentrator module, 16 channels |
|---|---|
| Inputs | 16 channels, each configurable: RTD, T/C, mV, resistance/potentiometer |
| RTD | 2-, 3-, 4-wire; Pt, Cu, Ni |
| Thermocouples | J, K, E, T, R, S, N, B, C |
| Resistance/potentiometer | 0-4000 ohms |
| Millivolt | -50 to 1000mV |
| Output | HART digital only |
| Power | 15-30DC, intrinsically safe |
| Housing | DIN rail mount, DIN rail enclosure |
| Configuration | PC, USB cable |
| Integral display | No |
| Alarm / relay output | No |
| Hazardous area certifications | Non-Incendive/Div. 2, Zone 2, Intrinsically-Safe |
| For use in SIS | No |
| RoHS | Yes |
Part numbers
Show all part numbers (1)
TCM/TPRG/HART/15-30DC [DIN]
How it compares with similar models
| Parameter | TCM temperature concentrator module, 16 channels, HART, 15-30DC, intrinsically safe, DIN rail | THZ3 dual input HART temperature transmitter for DIN rail, 4-20mA, 12-42DC | TRY PC-programmable isolated temperature transmitter, 4-20mA, 10-42DC, DIN rail | THZ3 associated IS dual input HART temperature transmitter (-AIS), 4-20mA, DIN rail |
|---|---|---|---|---|
| Type | Temperature concentrator module, 16 channels | Dual input HART temperature transmitter for DIN rail mounting | PC-programmable isolated temperature transmitter | Dual input associated IS HART temperature transmitter |
| Output | HART digital only | 4-20mA, isolated, linear, with HART | 4-20mA, linear, scalable, 4-20mA only | 4-20mA with HART |
| Power | 15-30DC, intrinsically safe | 12-42DC, 2-wire, loop powered | 10-42DC, 2-wire, loop powered | 12-42DC, 2-wire, loop powered |
| Configuration | PC, USB cable | PC, USB cable, HART | PC, USB cable | PC, USB cable, HART |
| Hazardous area certifications | Non-Incendive/Div. 2, Zone 2, Intrinsically-Safe | Non-Incendive/Div. 2, Zone 2 | — | Non-Incendive/Div. 2, Zone 2, Associated Intrinsically-Safe |
Request a price and lead time
TCM temperature concentrator module, 16 channels, HART, 15-30DC, intrinsically safe, DIN rail. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which signals can a TCM channel accept?
RTD, T/C, millivolt or resistance/potentiometer. Each of the 16 channels is configured individually.
How can the data reach a MODBUS or Ethernet host?
Over MODBUS RTU through the HMC HART-to-MODBUS converter, or over MODBUS/TCP and HART-IP through the HES HART-to-Ethernet Gateway System.
Where can the TCM be installed?
In Class I, II, III, Division 1 (Zone 0/1) hazardous areas, thanks to the intrinsically safe design. One isolated I.S. barrier serves all 16 points.
Similar equipment from other manufacturers