UNIMET® 300ST electrical safety test system for medical beds and equipment
The UNIMET® 300ST is a compact, lightweight tester for hospital and healthcare beds and for medical electrical equipment without patient connections. It runs automatic or manual test sequences, stores 600 data records and offers RS-232, PS/2 and Centronics interfaces.
- Type
- Electrical safety test system
- Objects tested
- Hospital beds, medical electrical equipment without patient connections
- Test types
- Visual, functional, electrical
- Measurements
- Insulation resistance, PE resistance, leakage current
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
Hospital beds, healthcare beds and medical electrical equipment without patient connections are the target of the UNIMET® 300ST. The system is compact and light, and it serves for periodic checks under DGUV regulation 3 in hospitals and healthcare facilities.
Tests run as an automatic or a manual sequence made up of a visual inspection, a functional test and electrical measurements. The measurements cover insulation resistance, PE resistance and leakage current. Every result is classified as passed or failed and written to a memory of 600 data records.
Operation uses four keys. A backlit LCD shows 4 lines of 20 characters. The tester has RS-232, PS/2 and Centronics interfaces, so a printer, a keyboard and a barcode scanner can be connected. UNIData 300 software is compatible for handling the stored results.
The test sequences follow IEC 62353, DIN EN 62353 (VDE 0751-1) and DIN VDE 0701-0702. Two versions are listed: UNIMET® 300ST (D/D), part number B96023000, and UNIMET® 300ST (CH/CH), part number B96023001.
Options include a barcode scanner with PS/2 connection (B96020082) and the DS32A and DS32 DCT three-phase adaptors. The calibration interval is 36 months, and the test probe is calibrated from the device settings menu.
Technical specifications
| Type | Electrical safety test system |
|---|---|
| Objects tested | Hospital beds, medical electrical equipment without patient connections |
| Test types | Visual, functional, electrical |
| Measurements | Insulation resistance, PE resistance, leakage current |
| Memory | 600 data records |
| Display | Backlit LCD, 4 lines of 20 characters |
| Interfaces | RS-232, PS/2, Centronics |
| Calibration interval | 36 months |
| Software version | 3.2 and later |
| Standards | IEC 62353, DIN EN 62353 (VDE 0751-1), DIN VDE 0701-0702 |
Part numbers
Show all part numbers (5)
B96023000 (D/D), B96023001 (CH/CH), B96020082, B96020098, B96020100
How it compares with similar models
| Parameter | UNIMET® 300ST electrical safety test system for medical beds and equipment | UNIMET® 610ST test system for the electrical safety of equipment and machines | UNIMET® 400ST safety tester for medical electrical equipment to IEC 62353 | STS 5000 Multifunction Test System, 800 A AC and 400 A DC Outputs |
|---|---|---|---|---|
| Interfaces | RS-232, PS/2, Centronics | RS-232, USB, PS/2, Ethernet | RS-232, PS/2, Centronics | USB, Ethernet, IEC 61850-9-2 |
| Software version | 3.2 and later | from 4.3.0 | from 3.2 | — |
Request a price and lead time
UNIMET® 300ST electrical safety test system for medical beds and equipment. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which measurements does the UNIMET® 300ST make?
Insulation resistance, PE resistance and leakage current, together with visual and functional tests. Each result is marked as passed or failed.
How many results can it store?
The memory holds 600 data records. The tester has an RS-232 interface, and UNIData 300 software is compatible for processing results.
Which versions are listed?
UNIMET® 300ST (D/D), part number B96023000, and UNIMET® 300ST (CH/CH), part number B96023001.
Similar equipment from other manufacturers