UNIMET® 400ST safety tester for medical electrical equipment to IEC 62353
The UNIMET® 400ST measures protective earth conductor resistance, insulation resistance, leakage currents, touch current and mains voltage to IEC 62353 and DIN VDE 0701-0702. It stores 600 data records and has RS-232, PS/2 and Centronics interfaces.
- Device type
- medical electrical equipment safety tester
- Test parameters
- protective earth conductor resistance; insulation resistance; PE current; equipment leakage current; touch current; applied part leakage current; mains voltage; current consumption
- Test standards
- IEC 62353:2007-05; DIN EN 62353 (VDE 0751-1):2008-08; ÖVE/ÖNORM EN 62353:2009-01; DIN VDE 0701-0702:2008-06; ÖVE E8701-1:03-01
- Interfaces
- RS-232, PS/2, Centronics
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The UNIMET® 400ST checks the electrical safety of medical electrical equipment. It performs the measurements required by IEC 62353:2007-05 and DIN EN 62353 (VDE 0751-1):2008-08, and also by ÖVE/ÖNORM EN 62353:2009-01, DIN VDE 0701-0702:2008-06 and ÖVE E8701-1:03-01.
The test parameters are protective earth conductor resistance, insulation resistance, PE current, equipment leakage current, touch current, applied part leakage current, mains voltage and current consumption. The tester works with a PAT Box for connecting the items under test and with a barcode scanner on a PS/2 connection.
The tester can be operated at the device itself, from a keyboard, or by reading a barcode with the scanner. A printer and a keyboard can be connected. The memory holds 600 data records, and the interfaces are RS-232, PS/2 and Centronics. The software is version 3.2 or later, and the settings include a probe adjustment.
The calibration interval is 36 months. The D/D version (part number B96024000) and the GB/GB version (part number B96024001) differ in the mains plug and socket type.
Technical specifications
| Device type | medical electrical equipment safety tester |
|---|---|
| Test parameters | protective earth conductor resistance; insulation resistance; PE current; equipment leakage current; touch current; applied part leakage current; mains voltage; current consumption |
| Test standards | IEC 62353:2007-05; DIN EN 62353 (VDE 0751-1):2008-08; ÖVE/ÖNORM EN 62353:2009-01; DIN VDE 0701-0702:2008-06; ÖVE E8701-1:03-01 |
| Interfaces | RS-232, PS/2, Centronics |
| Data storage | 600 data records |
| Calibration interval | 36 months |
| Software version | from 3.2 |
| Country versions | D/D; GB/GB |
Part numbers
Show all part numbers (4)
B96024000 (D/D), B96024001 (GB/GB), B96020082 (barcode scanner PS/2), B96020096 (PAT Box)
How it compares with similar models
| Parameter | UNIMET® 400ST safety tester for medical electrical equipment to IEC 62353 | UNIMET® 810ST safety tester for medical and industrial equipment | UNIMET® 610ST test system for the electrical safety of equipment and machines | UNIMET® 300ST electrical safety test system for medical beds and equipment |
|---|---|---|---|---|
| Device type | medical electrical equipment safety tester | safety tester for medical and industrial equipment | test system for electrical equipment and machines | — |
| Interfaces | RS-232, PS/2, Centronics | RS-232, USB, PS/2, Ethernet (optional) | RS-232, USB, PS/2, Ethernet | RS-232, PS/2, Centronics |
| Software version | from 3.2 | — | from 4.3.0 | 3.2 and later |
Request a price and lead time
UNIMET® 400ST safety tester for medical electrical equipment to IEC 62353. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which parameters does the tester measure?
Protective earth conductor resistance, insulation resistance, PE current, equipment leakage current, touch current, applied part leakage current, mains voltage and current consumption.
Which standards apply to the tests?
IEC 62353:2007-05, DIN EN 62353 (VDE 0751-1):2008-08, ÖVE/ÖNORM EN 62353:2009-01, DIN VDE 0701-0702:2008-06 and ÖVE E8701-1:03-01.
Which interfaces and how much memory?
RS-232, PS/2 and Centronics interfaces, and a memory for 600 data records.