The TDX 5000 carries out high voltage capacitance, tan delta (dissipation factor) and power factor measurements in a compact standalone housing. With the optional RCTD inductor, it also tests rotating machines.

Maximum output voltage
12000 V AC
Frequency
1 to 500 Hz
Output current
up to 300 mA (240 s); 100 mA continuous
Capacitance
0 pF to 5 µF; 5 µF to 200 µF

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The instrument is built as a self-contained unit with its own colour display, memory and communication ports, so no separate base unit is needed. Its housing measures 530 mm high, 450 mm wide and 215 mm deep and weighs 39 kg. It runs from a 100-230 V supply at 50-60 Hz with a maximum current of 16 A, and normal power draw stays below 1 kW. Operating temperatures range from -10 to 55 °C, storage from -20 to 70 °C, and the unit tolerates relative humidity from 0 to 95% without condensation.

Capacitance is measured across two ranges, from 0 pF to 5 µF and from 5 µF to 200 µF, while tan delta can be read on three ranges: 0 to 10% for capacitive measurements, 0 to 100%, and above 100%. Typical accuracy for tan delta and power factor is 0.05% of the value ± 0.005%, and the maximum error is 0.1% of the value ± 0.005%. Up to 64 test plans can be stored, along with more than 1000 results.

The high voltage generator delivers a variable AC output of up to 12000 V at a frequency adjustable from 1 to 500 Hz. Output current reaches 300 mA for up to 240 s, and it can be held at 100 mA continuously at 50/60 Hz. For objects with a large capacitive load, the optional RCTD inductor, rated at 40 H, raises the test current so that full voltage can be reached, carrying 0.6 A continuously and 1 A at most.

The test set is used to diagnose the insulation of power transformers, current and voltage transformers, bushings, circuit breakers, capacitor banks, motors and generators, and it can also check insulating oil. Besides capacitance and tan delta, it measures the magnetising current of a transformer. Tests run fully automatically, the PADS software handles evaluation and report generation, and a computer connects through USB or Ethernet. Testing methods follow IEC 61869-2, IEC 61869-3, EN 60076-1, ANSI/IEEE C57.12.90 and IEEE 62-1995, and electrical safety is covered by CEI EN 61010-1.

Technical specifications

Maximum output voltage12000 V AC
Frequency1 to 500 Hz
Output currentup to 300 mA (240 s); 100 mA continuous
Capacitance0 pF to 5 µF; 5 µF to 200 µF
Tan δ0 to 10% (capacitive); 0 to 100%; above 100%
Tan δ and power factor error, typical0.05% of the value ± 0.005%
Tan δ and power factor error, maximum0.1% of the value ± 0.005%
DisplayTFT, 640x480, colour, 132x99 mm
Memoryup to 64 test plans, over 1000 results
InterfacesUSB, Ethernet
Supply100-230 V ± 15%, 50-60 Hz; up to 16 A
Dimensions530 × 450 × 215 mm
Weight39 kg
Operating temperature-10 to 55 °C
Storage temperature-20 to 70 °C
RCTD inductor (option)40 H, 0.6 A continuous, 1 A maximum

Part numbers

Show all part numbers (10)

65175, PII66175, PII67175, PII10176T, PII42175, PII43175, PII47175, PII40175, PII13175, PII44175

Request a price and lead time

TDX 5000 standalone tan delta and capacitance test instrument, up to 12 kV. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Does the TDX 5000 need a separate base unit to operate?

No, the TDX 5000 is a compact standalone instrument with its own display, memory and USB and Ethernet interfaces.

How are large generators and motors with high capacitance tested?

The optional RCTD inductor, rated at 40 H, raises the test current so that full voltage can be reached on such objects.

What tan delta accuracy can be expected?

Typical error is 0.05% of the value ± 0.005%, and the maximum is 0.1% of the value ± 0.005%.

Similar equipment from other manufacturers

LET’S CONNECT

Let’s start with your challenge.

Complete a short brief. We will prepare an email you can send from your own email app.

Call us044 339-50-2009:00–18:00
USK ENERGY / SEARCH

Search

Solutions, specialist directions and guides