TD 5000 High-Voltage Module for Tan Delta and Capacitance Measurement, 12 kV

The TD 5000 extends the STS system, measuring loss tangent, power factor and capacitance of transformers and bushings. Measurements are taken at mains frequency or across a frequency range from 1 to 500 Hz.

Output voltage
from 12 V to 12 kV
Output frequency
1 to 500 Hz
Output current at 12 kV
100 mA continuous; 125 mA over 1 h; 300 mA for 240 s
Loss tangent
0 to over 100 %

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The module connects to the STS 5000 or STS 4000 instrument, which controls the test, powers the module and displays the results. Together they form a set for diagnosing the insulation of power transformers, bushings, current and voltage transformers, circuit breakers and rotating machines. Besides loss tangent, the module measures the magnetising current of a transformer, its inductance and its impedance.

The high-voltage generator uses electronic control and provides voltage from 12 V to 12 kV. At 12 kV the output current is 100 mA continuously, 125 mA for more than one hour and 300 mA for 240 s. Loss tangent is measured from 0 to over 100 %, power factor from 0 to 100 %, and capacitance from 1 pF to 200 µF in two sub-ranges, from 1 pF to 5 µF and from 5 µF to 200 µF. Magnetising current is measured on the 10 mA and 300 mA sub-ranges, and impedance from 1 kOhm to 1400 MOhm.

Several accessories work with the module. The CAP CAL reference capacitor, rated 200 pF, serves for calibration. The Stoil Cell, a 12 kV cell with a volume of 1 l, checks insulating oil, and the RCTD reactor is used for measurements on motors and generators with large insulation capacitance. A digital thermometer and hygrometer, with a range from -10 to +60 °C and from 5 to 95 % RH, records temperature and humidity during a test.

The module measures 440 × 345 × 210 mm and weighs 25 kg. Results are processed, assessed and reported in the PADS software. It comes with test cables and a folding trolley, and an aluminium transport case is available as an option. Test methods follow IEC 61869-2, IEC 61869-3, EN 60076-1, ANSI/IEEE C57.12.90 and IEEE 62-1995, safety meets CEI EN 61010-1, and electromagnetic compatibility meets EN 61326-1.

Technical specifications

Output voltagefrom 12 V to 12 kV
Output frequency1 to 500 Hz
Output current at 12 kV100 mA continuous; 125 mA over 1 h; 300 mA for 240 s
Loss tangent0 to over 100 %
Power factor0 to 100 %
Capacitance1 pF to 200 µF
Magnetising current10 mA and 300 mA ranges
Impedance1 kOhm to 1400 MOhm
Generatorhigh-voltage, electronically controlled
Dimensions440 × 345 × 210 mm
Weight25 kg
SoftwarePADS
CAP CAL reference capacitor200 pF
CompatibilitySTS 5000, STS 4000

Part numbers

Show all part numbers (7)

11175, PII11175, 39175, PII40175, PII13175, PII44175, PII47175

Request a price and lead time

TD 5000 High-Voltage Module for Tan Delta and Capacitance Measurement, 12 kV. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Does the TD 5000 work on its own?

No. The module adds to the STS series instrument (STS 5000 or STS 4000), which controls the test and shows the result.

How long can the module run at 12 kV?

At 100 mA it runs continuously, at 125 mA for more than one hour and at 300 mA for up to 240 s.

What is needed for measurements on rotating machines?

For motors and generators with large insulation capacitance, the module is supplemented with an RCTD reactor.

Similar equipment from other manufacturers