Stoil Cell for Insulating Oil Testing, Voltages up to 12 kV

Stoil Cell is a high-voltage test cell for dielectric oil. The result shows whether the insulating properties of the oil meet requirements and whether the oil is contaminated.

Purpose
dielectric oil testing
Measured parameter
oil loss tangent
Maximum voltage
12 kV
Volume
about 1 l

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The cell is auxiliary equipment for diagnosing power transformers. A sample of insulating oil is poured into the cell, a test voltage is applied, and the loss tangent of the oil is determined. This value shows whether the insulating characteristics of the liquid are within limits and whether contaminants that weaken its dielectric properties are present.

The cell is a glass vessel with built-in electrodes. Its volume is about 1 l, the empty cell capacitance is 60 pF, and the maximum test voltage is 12 kV.

The cell works as an option to the TD 5000 high-voltage module. Its electrodes are connected to an output of the module, which generates the test voltage and performs the measurement. The result, the oil loss tangent, is displayed on the STS 5000 instrument. The cell has no power source and no indicator of its own, so an STS 5000 with a TD 5000 module is required to use it.

Typical use is checking the oil of power transformers and monitoring its condition and contamination. The same STS 5000 instrument with a TD 5000 module also measures the capacitance and loss tangent of the transformer itself and its bushings, so oil testing can be carried out with the same set on the same site.

The cell is ordered under code 13175 as an additional item to a set with a TD 5000 module. The TD 5000 measures loss tangent at mains frequency or across a range of frequencies, so oil is tested by the same method used for transformer insulation measurement.

Technical specifications

Purposedielectric oil testing
Measured parameteroil loss tangent
Maximum voltage12 kV
Volumeabout 1 l
Empty cell capacitance60 pF
Constructionglass vessel with built-in electrodes
Connectionto TD 5000 module
Result displaySTS 5000 instrument

Part numbers

Show all part numbers (1)

13175

Request a price and lead time

Stoil Cell for Insulating Oil Testing, Voltages up to 12 kV. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Which equipment does the cell work with?

Its electrodes connect to the TD 5000 module, and the result is shown on the STS 5000 instrument.

What volume of oil does the test need?

The cell has a volume of about 1 l, and its empty capacitance is 60 pF.

What is the maximum test voltage for oil in the cell?

The maximum test voltage of the cell is 12 kV.