International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

Myrkos Lab Package — laboratory DGA package for dissolved gas analysis, 9 gases

The package combines the Myrkos micro-chromatograph for gases dissolved in oil with a syringe shaker. It measures 9 gases in 2 minutes and passes results to PPMreport for transformer diagnostics.

Type
Laboratory DGA package with micro GC and syringe shaker
Analysis time
2 minutes
Analysed gases
H₂, CO, CO₂, CH₄, C₂H₂, C₂H₄, C₂H₆, O₂, N₂
Sampling
ShakeTest syringe, 100 mL, for liquid; glass syringe, 10 mL, for gas from relay or headspace

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The Myrkos Lab Package is designed for laboratories that analyse gases dissolved in the insulating liquid of power and distribution transformers. It increases the number of samples a laboratory can process in a shift: samples are prepared by the automated Shake Test on a syringe shaker, after which the headspace gas is fed to the micro-chromatograph. Results are produced as a one-page DGA report in PDF, and the data can be exported to third-party software for further diagnosis.

The analyser works on the principle of gas chromatography: sample components are separated in columns and recorded by a detector, and one sample takes 2 minutes to measure. Hydrogen H₂, carbon monoxide CO, carbon dioxide CO₂, methane CH₄, acetylene C₂H₂, ethylene C₂H₄, ethane C₂H₆, oxygen O₂ and nitrogen N₂ are determined. Helium of 99.999 % purity at 80 psig serves as the carrier gas, and the carrier gas module runs for 42 hours. Calibration is monitored with a 920 cc TALL CALGAS cylinder traceable to NIST.

Liquid samples are taken with a 100 mL ShakeTest syringe, and gas samples from a gas relay or the headspace with a 10 mL glass syringe. The package works with mineral oil, silicone fluid and natural and synthetic esters.

The package includes PPMreport control software, MyrkosView basic DGA analysis software and, as an option, the InsideView diagnostic program. A web application, Oil LIMS, can be ordered separately to store results and keep transformer histories. Connection to the computer is by TCP/IP directly or over a network, and the package's procedures can be built into the laboratory's quality management system.

The hardware is based on the field version, the Myrkos 2, housed in a waterproof wheeled case with a telescopic handle. It operates from 0 to 50 °C, can be stored from 0 to 75 °C, tolerates 5–95 % humidity without condensation and works up to 4500 m. The enclosure is rated IP67 closed and IP20 open under IEC 60529. The method follows ASTM D3612-02 and IEC 60567, with electrical safety to IEC/EN 61010-1 and EMC to IEC 61326-1 and EN 55011, and the case passed a 1 m drop test to ISTA 3A.

Technical specifications

TypeLaboratory DGA package with micro GC and syringe shaker
Analysis time2 minutes
Analysed gasesH₂, CO, CO₂, CH₄, C₂H₂, C₂H₄, C₂H₆, O₂, N₂
SamplingShakeTest syringe, 100 mL, for liquid; glass syringe, 10 mL, for gas from relay or headspace
LiquidsMineral oil, silicone fluid, natural and synthetic esters
Carrier gasHelium 99.999 %, 80 psig
Helium module autonomy42 hours
Calibration gasTALL CALGAS 920 cc, NIST traceable
SoftwarePPMreport, MyrkosView, InsideView (optional)
ConnectivityTCP/IP direct or networked
Operating temperature0 to 50 °C
Storage temperature0 to 75 °C
Humidity5–95 % non-condensing
Altitude0–4500 m
Ingress protectionIP67 when closed, IP20 when open (IEC 60529)
Shipping weight42 kg
Standards metIEC 60567, ASTM D3612-02, IEC/EN 61010-1 (safety), IEC 61326-1 (EMC), EN 55011, ISTA 3A

Request a price and lead time

Myrkos Lab Package — laboratory DGA package for dissolved gas analysis, 9 gases. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Which gases does the package determine?

Nine gases: H₂, CO, CO₂, CH₄, C₂H₂, C₂H₄, C₂H₆, O₂ and N₂. One sample is measured in 2 minutes.

Which insulating liquids can the analyser handle?

Mineral oil, silicone fluid, and natural and synthetic esters.

How long does the carrier gas last?

The helium module of 99.999 % purity is rated for 42 hours of operation.

Similar equipment from other manufacturers

LET’S CONNECT

Let’s start with your challenge.

Complete a short brief. We will prepare an email you can send from your own email app.

Call us044 339-50-2009:00–18:00