DVD100, DVD200 and DVD300 resistive-capacitive voltage dividers, AC/DC, 100–300 kV

Modular resistive-capacitive voltage dividers for measuring high AC and DC voltages and for calibrating high-voltage measuring systems on site. Standard models for 100, 200 and 300 kV have a 10000:1 ratio, and their calibration is traceable to NIST standards.

Type
resistive-capacitive voltage divider
Voltage
50 – 300 kV AC/DC (standard models)
Division ratio
10000:1
Measurement functions
AC Average, AC RMS, AC Peak, AC Peak/√2, DC Average, DC Peak, DC Ripple

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The divider serves as the primary calibration reference for checking the accuracy of ultra-high-voltage AC and DC measuring systems on site. It is also used for general high-voltage measurement in the laboratory and in the field. Other instruments, such as an oscilloscope or a multimeter, can be connected to the divider.

This is a resistive-capacitive divider built to IEC 60060-2 and IEEE Std 4. The kit includes the high-voltage divider, the measuring section and connecting cables, all kept in a compact, rugged case. The KVM measuring head works with the 10000:1 divider and reads AC Average, AC RMS, AC Peak, AC Peak/√2, DC Average, DC Peak and DC Ripple. Readings appear on a 4½-digit LCD, with a range switch for the lower and upper ranges, and indicators for mains power and low battery.

The design is modular, so the divider comes apart quickly for transport while keeping its stability and small ratio error. The largest part is under 900 mm long and weighs less than 35 kg. Transport cases are an option, and other ratios can be built to order. The unit runs from mains or from a nickel-metal hydride (Ni-MH) battery.

The standard models are the DVD100 at 100 kV, the DVD200 at 200 kV and the DVD300 at 300 kV, all for AC and DC with a 10000:1 ratio. The range covers 50 to 300 kV, and higher voltages are available on request. The KVM measuring head is also offered for mounting in a 19-inch rack. Operating conditions are 0 to 40 °C indoors or outdoors in dry weather, humidity below 95% without condensation, and altitude up to 1000 m. Storage is from 0 to 55 °C in a dry room.

Technical specifications

Typeresistive-capacitive voltage divider
Voltage50 – 300 kV AC/DC (standard models)
Division ratio10000:1
Measurement functionsAC Average, AC RMS, AC Peak, AC Peak/√2, DC Average, DC Peak, DC Ripple
DisplayLCD, 4½ digits
Power supplymains or Ni-MH battery
Largest partlength under 900 mm, weight under 35 kg
Operating temperature0 – 40 °C
Humidityunder 95%, non-condensing
Altitudeunder 1000 m
Storage temperature0 – 55 °C
StandardsIEC 60060-2, IEEE Std 4

Part numbers

Show all part numbers (3)

DVD100, DVD200, DVD300

Request a price and lead time

DVD100, DVD200 and DVD300 resistive-capacitive voltage dividers, AC/DC, 100–300 kV. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Can the divider be used with an oscilloscope?

Yes. Apart from the KVM head, other instruments such as an oscilloscope or a multimeter can be connected to the divider.

How is the divider transported to site?

It comes apart into modules, and the largest part is under 900 mm long and under 35 kg. Transport cases are ordered as an option.

Can other ratios or higher voltages be supplied?

Yes, other ratios and voltages above 300 kV are built to order. The standard models have a 10000:1 ratio.