Derivation Box 3CH for connecting partial discharge instruments to sensors
The Derivation Box 3CH links a partial discharge measuring instrument, either permanently or temporarily, to sensors fitted to electrical equipment. Measurements can be made without taking the equipment out of service.
- Number of channels
- 3 (CH1, CH2, CH3) plus SYNC
- Transient voltage at CH outputs
- below 90 Vpeak
- Output voltage CH
- up to 50 Vrms
- Transient voltage at PH inputs
- up to 350 Vpeak
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
Surge arrestors inside the box limit the transient voltage at the outputs, which protects operators and the connected instrument. Transients on channels CH1, CH2 and CH3 are held below 90 Vpeak, and the maximum output voltage of each channel is 50 Vrms.
The PH1, PH2 and PH3 inputs withstand transient voltages up to 350 Vpeak. A separate SYNC output carries the reference voltage signal needed to phase-align partial discharge pulses. It operates at 50 to 60 Hz, with a maximum output voltage of 200 Vrms.
The measurement channels cover 100 kHz to 50 MHz, and every input and output connector is a female BNC type. The box can be connected either permanently or temporarily, as the job requires.
The enclosure is a thick powder-coated aluminium box. The standard indoor version is rated IP44, which rises to IP65 with an external protection box. An outdoor IP65 version is available as an option, and higher ratings up to IP68 can also be supplied. The operating range is -25°C to +65°C.
Installation in most cases does not require taking the electrical equipment out of service. The box is fixed to a support with the screws on its back panel, and wall mounting is preferred. A dedicated binding post is provided for earthing, and the earth cable is chosen for its length and local safety requirements. Matching accessories from the same range include the FMC flexible magnetic coupler, the HFCT high-frequency current transformer clamp and the PD calibrator.
Technical specifications
| Number of channels | 3 (CH1, CH2, CH3) plus SYNC |
|---|---|
| Transient voltage at CH outputs | below 90 Vpeak |
| Output voltage CH | up to 50 Vrms |
| Transient voltage at PH inputs | up to 350 Vpeak |
| Output voltage SYNC | up to 200 Vrms |
| SYNC channel frequency | 50–60 Hz |
| CH channel frequency | 100 kHz–50 MHz |
| Operating temperature | -25 °C to +65 °C |
| Degree of protection | IP44 (IP65 with external protection box), optionally up to IP68 |
| Connectors | BNC (female) |
| Housing | powder-coated aluminium |
| Mounting | wall, screws on the back panel |
Request a price and lead time
Derivation Box 3CH for connecting partial discharge instruments to sensors. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Does the equipment have to be shut down to install the box?
In most cases no. The box is connected to sensors that are already installed, so the equipment stays in service, and partial discharge measurements can also be made under voltage.
Can the box be installed outdoors?
The standard IP44 version is for indoor use. An outdoor IP65 version is available as an option, and protection ratings up to IP68 can also be supplied.
What is the SYNC output for?
It supplies the 50 to 60 Hz reference voltage signal used to phase-align partial discharge pulses, with a maximum output voltage of 200 Vrms.
Similar equipment from other manufacturers