TMZ MODBUS temperature transmitter and signal converter, dual universal input, RS-485, 24DC
The TMZ dual input model takes current, voltage, RTD, thermocouple, mV, potentiometer or resistance signals and converts them to MODBUS RTU (RS-485). Up to 32 units share one link without repeaters.
- Type
- PC-programmable MODBUS temperature transmitter and signal converter with dual universal input
- Input
- Two inputs, each: current, voltage, RTD, T/C, mV, ohms, potentiometer
- Output
- MODBUS RTU (RS-485)
- Power
- 24DC, 4-wire, line/mains powered
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
The Dual Universal Input (2PRG) model of the TMZ accepts two inputs, and each input can be configured for current, voltage, RTD, thermocouple, mV, potentiometer or resistance. This doubles the monitoring capability of a single unit. The inputs are converted to the standard MODBUS RTU (RS-485) protocol, ready for direct interface with MODBUS-based monitoring and control systems.
The unit is 4-wire and line/mains powered at 24DC. Up to 32 TMZs can be multidropped without repeaters onto a single low-cost link such as a twisted wire pair, so there is no need to run a dedicated wire for each signal. This cuts installation, cable, conduit, connection and wire tray costs. The TMZ can share a data link with standard MODBUS RTU devices, and any combination of TMZ configurations may be used in the same system.
Input resolution is 20-bit, giving accuracy of ±0.1°C (±0.18°F) with all platinum RTDs, or ±0.01% of maximum span with current and voltage inputs. The measurement reaches the control system as a digital signal, so the output error of a traditional analogue transmitter is eliminated. Up to 5 years are allowed between scheduled calibrations. Isolation and RFI/EMI protection reduce the effects of ground loops, plant noise, and radio frequency and electromagnetic interference.
All parameters are set on a single screen of the Windows software: input type and range, input trimming, MODBUS parameters, T/C reference junction compensation and standard or custom linearisation curves. The TMZ can be trimmed with two data points within the selected range, which concentrates accuracy on a critical segment. It is housed for DIN rail with no display, is CE conformant to EMC Directive 2014/30/EU, EN 61326, and is RoHS compliant. It is not for use in SIS.
Technical specifications
| Type | PC-programmable MODBUS temperature transmitter and signal converter with dual universal input |
|---|---|
| Input | Two inputs, each: current, voltage, RTD, T/C, mV, ohms, potentiometer |
| Output | MODBUS RTU (RS-485) |
| Power | 24DC, 4-wire, line/mains powered |
| Housing | DIN rail mount, DIN rail enclosure |
| Units per link | Up to 32 TMZs without repeaters |
| Input resolution | 20-bit |
| Accuracy | ±0.1°C (±0.18°F) with all platinum RTDs, ±0.01% of maximum span with current and voltage |
| Stability | Up to 5 years between scheduled calibrations |
| Configuration | PC, USB cable |
| Isolation | Fully isolated |
| Integral display | No |
| Intrinsically safe input connections | No |
| For use in SIS | No |
| Conformity | CE: EMC Directive 2014/30/EU, EN 61326; RoHS |
Part numbers
Show all part numbers (1)
TMZ/2PRG/MB/24DC
How it compares with similar models
| Parameter | TMZ MODBUS temperature transmitter and signal converter, dual universal input, RS-485, 24DC | TMZ MODBUS temperature transmitter and signal converter for RTD, T/C, mV, ohms and potentiometer, 24DC | THZ3 dual input HART temperature transmitter for DIN rail, 4-20mA, 12-42DC | TMZ MODBUS converter for current and voltage signals, 0-50mA and 0-10V, 24DC |
|---|---|---|---|---|
| Type | PC-programmable MODBUS temperature transmitter and signal converter with dual universal input | PC-programmable MODBUS temperature transmitter and signal converter | Dual input HART temperature transmitter for DIN rail mounting | PC-programmable MODBUS signal converter for current and voltage |
| Input | Two inputs, each: current, voltage, RTD, T/C, mV, ohms, potentiometer | One input: RTD, T/C, mV, ohms, potentiometer | Dual sensor, programmable: RTD, T/C, mV, ohms, potentiometer | Programmable, any range within 0-50mA or 0-10V |
| Output | MODBUS RTU (RS-485) | MODBUS RTU | 4-20mA, isolated, linear, with HART | MODBUS RTU |
| Power | 24DC, 4-wire, line/mains powered | 24DC, 4-wire, line/mains powered | 12-42DC, 2-wire, loop powered | 24DC, 4-wire, line/mains powered |
| Configuration | PC, USB cable | PC, USB cable | PC, USB cable, HART | PC, USB cable |
Request a price and lead time
TMZ MODBUS temperature transmitter and signal converter, dual universal input, RS-485, 24DC. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Which signals does the TMZ 2PRG accept?
Two inputs, each configurable for current, voltage, RTD, thermocouple, mV, potentiometer or resistance.
How many TMZs can share one link?
Up to 32 without repeaters, for example on one twisted wire pair, so no dedicated wire per signal is needed.
What measurement accuracy can be expected?
±0.1°C (±0.18°F) with all platinum RTDs and ±0.01% of maximum span with current and voltage inputs. Input resolution is 20-bit.