International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

TST TRUE SURFACE Thermocouple for Flat Surfaces, rated to 200°C

A small surface-mounting thermocouple whose junction is shielded from passing air, so it reads the plate rather than the draught over it. Typical accuracy and repeatability fall between one and two percent of ΔT, with a 200°C rating.

Type
TRUE SURFACE (TST) surface thermocouple
Operating temperature
up to 200°C (400°F)
Accuracy and repeatability
1–2% ΔT, typical
Measuring junction
isothermal block

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

Washer-style thermocouples on a hot plate tend to read low as soon as air moves across them. The TST tackles this by setting its measuring junction in an isothermal block with high thermal conductivity, then covering that block with a moulded insulator that keeps ambient airflow away.

The manufacturer compared the two designs on a brass hot plate. Each sensor was allowed to reach equilibrium, then room-temperature air was blown over the plate, with plate-to-ambient differentials of 20, 40, 60 and 80°C and an ambient of 25°C. Readings were taken again once the system settled, and the TST kept its error within one to two percent of ΔT.

The package measures 11.88 mm on a side and 6.10 mm high, so it fits into corners and other cramped spots; where even less height is available, the moulded cover can be removed. Fixing uses commonly available screw types.

Suggested uses include semiconductor process chambers, platens, packaging machinery, cleaning equipment and food preparation equipment. Maximum surface temperature is 200°C (400°F).

Technical specifications

TypeTRUE SURFACE (TST) surface thermocouple
Operating temperatureup to 200°C (400°F)
Accuracy and repeatability1–2% ΔT, typical
Measuring junctionisothermal block
Insulatormoulded, removable
Body size11.88 mm side, 6.10 mm high
Fixingcommon screw types

Request a price and lead time

TST TRUE SURFACE Thermocouple for Flat Surfaces, rated to 200°C. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Why is the TST more accurate than a washer thermocouple?

Its junction sits in an isothermal block covered by a moulded insulator, so moving air has far less effect on the reading.

What is the maximum surface temperature?

200°C (400°F).

Will it fit in a tight corner?

The body is 11.88 mm on a side and 6.10 mm high, and the moulded cover can be taken off for an even smaller footprint.