International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

Series 4060 Process Gas Chromatograph with FID, PID, TCD or PED Detector

An on-line gas chromatograph for automatic impurity analysis in process and high-purity gases. Four detector technologies are available, and the analyser can be built into a standard 19 inch instrument rack or supplied as a wall-mounted unit.

Detectors
FID, PID, TCD, PED
Components
hydrocarbons, most gases, selected compounds
Concentration level
trace or percent (detector dependent)
Sensitivity
1 % FS

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

Choosing a 4060 starts with the detector. The 4060/FID uses flame ionisation and picks up hydrocarbons at trace or percent concentrations; the 4060/PID relies on photoionisation for certain hydrocarbons at trace level; the 4060/TCD measures thermal conductivity and suits the majority of gases; and the 4060/PED, a plasma emission instrument, targets selected compounds at trace level, nitrogen in hydrogen being one example.

Typical jobs include checking beverage-grade carbon dioxide for benzene, watching for aromatic contamination on air liquefaction and other gas production plant, tracing benzene and other pollutants in ambient air, and certifying gas purity or newly installed gas lines. Speciality gas laboratories, chemical works, speciality steelmakers, semiconductor fabs and gas management systems are the usual users.

Each injection is taken by a 10-port, two-position GC valve fitted with a loop of fixed volume, and separation happens on a packed column made of stainless steel. Sample chamber, detector and column are each held at temperature by their own controller, the column typically at 65 °C to 70 °C. Sensitivity is 1 % FS and accuracy ±2 % FS at constant temperature, with repeatability of 1 % FS on every range, zero and span noise under 0.5 % FS and drift of 2 % FS per week. Response time depends on the application.

Results appear on a red 5 digit LED and a vacuum fluorescent display. Outputs are 0-1 V DC plus an isolated 4-20 mA loop into up to 500 ohms; with more than 3 components one output serves as a gas ID signal. Form-C relays rated 3 A at 250 V AC handle one system alarm and two concentration alarms, and an RS-232 port is fitted.

The chromatograph runs on 110 or 220 V AC, 50/60 Hz, draws up to 600 VA and operates between 4 °C and 43 °C with a standard sample flow of 0.5 l/min. The rack version measures 483 x 221 x 623 mm, the wall version 762 x 762 x 328 mm.

Technical specifications

DetectorsFID, PID, TCD, PED
Componentshydrocarbons, most gases, selected compounds
Concentration leveltrace or percent (detector dependent)
Sensitivity1 % FS
Accuracy±2 % FS at constant temperature
Repeatability1 % FS
Zero and span drift2 % FS per week
Sampling valve10-port, 2-position, fixed-volume loop
Analogue outputs0-1 V DC; isolated 4-20 mA, up to 500 ohms
AlarmsForm-C 3 A 250 V AC: system + 2 concentration
InterfaceRS-232
Supply110 or 220 V AC, 50/60 Hz, up to 600 VA
Operating temperature4 °C to 43 °C
Dimensionsrack 483 x 221 x 623 mm; wall 762 x 762 x 328 mm

Request a price and lead time

Series 4060 Process Gas Chromatograph with FID, PID, TCD or PED Detector. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Which detector suits hydrocarbon analysis?

The 4060/FID covers hydrocarbons at trace or percent level, while the 4060/PID handles certain hydrocarbons at trace level. Most other gases are measured with the 4060/TCD.

Can it measure nitrogen in hydrogen?

Yes, the 4060/PED with plasma emission detection is configured for selected trace compounds, including nitrogen in hydrogen.

How does it pass results to a control system?

Through 0-1 V DC and isolated 4-20 mA outputs, Form-C relays rated 3 A at 250 V AC and an RS-232 serial port.

Similar equipment from other manufacturers

LET’S CONNECT

Let’s start with your challenge.

Complete a short brief. We will prepare an email you can send from your own email app.

Call us044 339-50-2009:00–18:00
USK ENERGY / SEARCH

Search

Solutions, specialist directions and guides