8801B gas moisture monitor with wall-mounted display unit, -100 to 20 °C dew point

The 8801B pairs a nanopore dew point sensor with a wall-mounted unit that powers it, shows the reading in large numerals and retransmits it on an isolated 4–20 mA output. Trend graphs and a histogram appear on the same screen.

Dew point range
-100 to 20 °C
Accuracy
±2 °C, temperature corrected
Repeatability
0.8 °C
Response time
95 % of step change in 3 min

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

Power comes in through a self-configuring supply that detects what it has been connected to: 90 to 260 V AC at 47 to 440 Hz as standard, or 12 to 24 V DC at 10 VA as an option. The isolated 4–20 mA output can sink or source. Alarm relays rated 3 A at 250 V AC or 30 V DC have programmable hysteresis. Options include isolated RS-422/485 with Modbus ASCII and RTU, rechargeable batteries for data logging during mains loss, and built-in self-verification and calibration.

Measurement performance matches the 8801 sensor family: -100 to 20 °C dew point, ±2 °C accuracy with temperature correction, 0.8 °C repeatability, and 95 % of a step change within 3 minutes at sample flows above 1 litre per minute. A separate temperature sensor reads -40 to 70 °C to ±2 °C.

The enclosure is impact-resistant ABS measuring 5.16″ × 7.32″ × 3.25″ and weighing 0.7 kg, rated IP65. A rear suspension bracket hides the fixings, and the terminal compartment is separate and pre-punched for cable glands. Shock-mount vibration isolation can be fitted where the wall or skid is not steady. Electronics are specified for -40 to 85 °C, operation for -20 to 70 °C and storage for -30 to 80 °C; insulation resistance exceeds 100 MΩ at 500 V DC.

Sites using it include natural gas treatment, petrochemical and industrial gas plants, power stations, air dryers and heat treatment furnaces, as well as pharmaceutical, aerospace and medical facilities. Calibration is NIST traceable, and a virtual analyser for the PC reproduces the instrument interface with spoken explanations, so operators can train before they touch the real unit.

Technical specifications

Dew point range-100 to 20 °C
Accuracy±2 °C, temperature corrected
Repeatability0.8 °C
Response time95 % of step change in 3 min
Sample flow> 1 l/min
Temperature sensor-40 to 70 °C, ±2 °C
Supply90–260 V AC, 47–440 Hz; optional 12–24 V DC, 10 VA
Outputsisolated 4–20 mA, relays 3 A 250 V AC / 30 V DC
Digital interfaceoptional RS-422/485, Modbus ASCII and RTU
Insulation resistance> 100 MΩ at 500 V DC
EnclosureABS, wall-mounted, 5.16″ × 7.32″ × 3.25″
ProtectionIP65
Weight0.7 kg
Operating temperature-20 to 70 °C
Storage temperature-30 to 80 °C

Request a price and lead time

8801B gas moisture monitor with wall-mounted display unit, -100 to 20 °C dew point. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

How does the 8801B differ from the 8801T transmitter?

The 8801B adds a wall-mounted display unit with mains supply, isolated 4–20 mA retransmission and 3 A alarm relays, whereas the 8801T is a bare loop-powered sensor.

Can the monitor run from a DC supply?

Yes, a 12–24 V DC input at 10 VA is available as an option, and the unit detects the supply type automatically.

Is it suitable for mounting on vibrating equipment?

The IP65 ABS enclosure can be ordered with shock-mount vibration isolation for that purpose.

Similar equipment from other manufacturers