6400TSG Total Sulphur Analyser for Fuel Gas, UV Fluorescence, Ex px Cabinet
An on-line system that burns a fuel-gas sample in a flame converter and reads the resulting SO2 by UV fluorescence, giving a continuous total sulphur figure. Analyser, converter and sample handling share one pressurised steel cabinet certified for hazardous areas.
- Measured value
- total sulphur in fuel gas
- Method
- UV fluorescence after flame converter
- Range
- 0–1 ppm (H2 only) or 0–10 ppm up to 0–500 ppm
- Repeatability
- ±2% of full scale
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
Each unit is set up and calibrated at the factory for a single range. The top range is 0–500 ppm total sulphur; the bottom is 0–1 ppm, but only when the stream is 100% hydrogen, otherwise the lowest available range is 0–10 ppm. Repeatability is ±2% of full scale, zero and span noise stay under 1%, drift is 2% of full scale per week, and the reading reaches 90% in less than 100 seconds at the standard 5 l/min sample flow.
The method follows ASTM D6667. Three main assemblies do the work: the 6400E UV fluorescence SO2 analyser, a flame combustion converter module (FCCM) and the sample conditioning. Switching between process sample and calibration gases is left to the operator's own control system, which the cabinet interfaces with.
The cabinet is marked Ex px IIC T3 Gb against EN 60079-0:2009 and EN 60079-2:2007, under certificate ITS11ATEX17425X. A floor-standing build with three certified viewing windows and a wall-hung build with one are covered; purging needs at least 340 SLPM of nitrogen or air from a 4 to 5.5 bar supply, for 6 minutes on the floor version and 10 minutes on the wall version, or 15 minutes with N2.
Utilities are sulphur-free UHP air for the converter at 40 psi and 1 l/min, dryer air at 25 psi and 3 l/min with a -40°C dew point, sulphur-free UHP hydrogen at 40 psi, purge gas at 40 psi, plus zero and span gases. Power is 110 or 220 V AC, 50/60 Hz, 500 W. The isolated 4–20 mA signal drives loads up to 500 ohms, and one system alarm plus two Form-C concentration relays rated 3 A at 125 V AC are fitted. Ambient operating range is 5°C to 40°C.
Technical specifications
| Measured value | total sulphur in fuel gas |
|---|---|
| Method | UV fluorescence after flame converter |
| Range | 0–1 ppm (H2 only) or 0–10 ppm up to 0–500 ppm |
| Repeatability | ±2% of full scale |
| Response time | 90% in under 100 s |
| Zero/span drift | 2% of full scale per week |
| Sample flow | 5 l/min |
| Output | 4–20 mA, isolated, up to 500 ohms |
| Alarms | 1 system + 2 Form-C relays, 3 A, 125 V AC |
| Power | 110 or 220 V AC, 50/60 Hz, 500 W |
| Hazardous-area marking | Ex px IIC T3 Gb |
| Purge | 340 SLPM minimum, 4–5.5 bar supply |
| Operating temperature | 5°C to 40°C |
| Compliance | ASTM D6667 |
Request a price and lead time
6400TSG Total Sulphur Analyser for Fuel Gas, UV Fluorescence, Ex px Cabinet. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
What is the lowest range available?
0–1 ppm total sulphur, but only for a 100% hydrogen stream; for other gases the lowest range is 0–10 ppm.
Can the cabinet go into a Zone 1 area?
It carries Ex px IIC T3 Gb marking under EN 60079-0 and EN 60079-2, so it is intended for gas atmospheres requiring Gb equipment.
Which support gases does it consume?
Sulphur-free UHP air for the converter, dryer air, sulphur-free UHP hydrogen, nitrogen or air for purging, and zero and span calibration gases.
Similar equipment from other manufacturers