International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

6400T UV fluorescence sulphur dioxide analyser, full scale selectable 0-50 ppb up to 0-20,000 ppb

A rack-mounted continuous analyser for trace sulphur dioxide, working on ultraviolet fluorescence. Sister builds measure hydrogen sulphide alongside SO2 (6400T-H2S) or cover higher SO2 concentrations (6400TH).

Measured gas
SO2; both SO2 and H2S on the 6400T-H2S
Method
UV fluorescence
Range
0-50 ppb up to 0-20,000 ppb
Units
ppb, ppm, µg/m3, mg/m3

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

Stability at parts-per-billion level is the point of this design. An optical shutter corrects for photomultiplier drift, while a reference detector follows any change in UV lamp output, so the zero holds to < 0.5 ppb over 24 hours and < 1 ppb over 7 days. A hydrocarbon kicker and the optical layout keep interfering compounds out of the reading.

On the base 6400T the user sets full scale between 0-50 ppb and 0-20,000 ppb, with dual ranges and auto-ranging available. Zero noise is < 0.2 ppb RMS, span noise < 0.5% of reading above 50 ppb and the lower detectable limit 0.4 ppb. Lag time is 20 seconds, rise and fall take < 100 seconds to 95%, linearity stays within 1% FS and precision 0.5% of reading above 50 ppb. Output units are ppb, ppm, µg/m3 or mg/m3, chosen by the operator.

For sulphide work the 6400T-H2S adds a catalytic converter held at 315°C that oxidises H2S to SO2 while leaving harder-to-oxidise sulphur compounds alone; a switching mode then reports H2S and SO2 from one instrument. The series also forms the basis of the 6400TSG total sulphur system with an FCCM converter, designed around ASTM D6667.

The chassis is 178 x 432 x 597 mm and weighs 16.2 kg, drawing sample at 650 cc/min ±10% on 100–120 V or 220–240 V, 50/60 Hz, between 5 and 40 °C. A colour touchscreen, internal data logging with programmable averaging, Ethernet, two RS232 ports, USB, up to four analogue outputs, eight digital outputs and six digital inputs are fitted, with an internal zero and span check as an option.

Technical specifications

Measured gasSO2; both SO2 and H2S on the 6400T-H2S
MethodUV fluorescence
Range0-50 ppb up to 0-20,000 ppb
Unitsppb, ppm, µg/m3, mg/m3
Zero noise< 0.2 ppb RMS
Lower detectable limit0.4 ppb
Zero drift< 0.5 ppb/24 h, < 1 ppb/7 days
Rise/fall time< 100 s to 95%
Linearity1% FS
Sample flow650 cc/min ±10%
Supply100–120 V, 220–240 V, 50/60 Hz
Operating temperature5 to 40 °C
Dimensions178 x 432 x 597 mm
Weight16.2 kg

Request a price and lead time

6400T UV fluorescence sulphur dioxide analyser, full scale selectable 0-50 ppb up to 0-20,000 ppb. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Can the 6400T measure hydrogen sulphide?

Yes, in the 6400T-H2S version a catalytic converter at 315°C turns H2S into SO2. Switching mode then gives both gases from one analyser.

What is the SO2 detection limit?

It detects down to 0.4 ppb, with zero noise below 0.2 ppb. The smallest full scale is 0-50 ppb.

How is data stored and retrieved?

The internal logger records concentrations and operating parameters with programmable averaging. Data is read from the front panel or over RS232, Ethernet or USB.