International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

3010MA Oxygen Purity Analyser for 95 to 100 % O2, Paramagnetic, 19-inch Rack

Bench and rack instrument for checking how pure an oxygen stream is, working only in the top few per cent of the scale. Typical duty is product gas from air separation plants and certification of medical-grade oxygen and air.

Measuring principle
paramagnetic
Spans
95–100 / 96–100 / 98–100 % O2, auto-ranging
Accuracy
±0.045 % O2 abs. at constant temperature; ±0.05 % O2 over operating range
Linearity
0.1 % O2

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

Most oxygen analysers spread their resolution over a wide scale. The 3010MA does the opposite: its three spans, 95–100, 96–100 or 98–100 % O2, sit right at the top, so a fraction of a per cent of contamination is easy to read. Switching between spans is automatic, and a separate range identification signal tells the control system which one is active.

Absolute accuracy is ±0.045 % O2 when ambient temperature is steady and ±0.05 % O2 across the permitted ambient band of 20 to 30 °C. Linearity is 0.1 % oxygen, repeatability ±0.10 %, and the reading gets to 90 % of its final value within 10 s at 25 °C.

Measurement relies on oxygen being strongly drawn into a magnetic field, far more than nitrogen, argon or other common gases. The sensing assembly is held in an oven at a constant 40 °C, and an 8-bit microprocessor runs the instrument.

Industries using it include air separation and liquefaction, bulk gas filling, medical gas qualification, inert blanketing checks, chemical reaction monitoring, semiconductor fabs and petrochemical plants.

Wiring to plant systems is straightforward: two 0–1 V DC and two isolated 4–20 mA outputs, an RS-232C port, two adjustable concentration alarms and a system failure contact. Mains supply is 90–130 or 200–240 V AC at 47/63 Hz. The unit fits a 19-inch rack or a panel cut-out, weighs 16 kg and takes 1/4 inch or 6 mm sample fittings. It is a general-purpose indoor instrument, not for hazardous areas, and is declared to EN55011 Class A, EN50082-2 and EN61010-1.

Technical specifications

Measuring principleparamagnetic
Spans95–100 / 96–100 / 98–100 % O2, auto-ranging
Accuracy±0.045 % O2 abs. at constant temperature; ±0.05 % O2 over operating range
Linearity0.1 % O2
Repeatability±0.10 % O2
Response time90 % of scale in < 10 s at 25 °C
Analogue outputs2 × 0–1 V DC, 2 × 4–20 mA isolated
Alarms2 concentration set points, system failure contact
InterfaceRS-232C
Supply90–130 or 200–240 V AC, 47/63 Hz
Operating temperature20–30 °C
Mounting19-inch rack or panel
Weight16 kg
Sample fittings1/4 inch or 6 mm

Request a price and lead time

3010MA Oxygen Purity Analyser for 95 to 100 % O2, Paramagnetic, 19-inch Rack. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Which spans can the 3010MA measure?

Three spans, 95–100, 96–100 or 98–100 % O2, with automatic switching and a range ID signal for the control system.

Is the analyser suitable for a hazardous area?

No. It is a general-purpose indoor instrument for non-hazardous locations with ambient temperature between 20 and 30 °C.

What signals are available for the DCS?

Two 0–1 V DC outputs, two isolated 4–20 mA outputs and an RS-232C port, plus two concentration alarms and a system failure contact.

Similar equipment from other manufacturers

LET’S CONNECT

Let’s start with your challenge.

Complete a short brief. We will prepare an email you can send from your own email app.

Call us044 339-50-2009:00–18:00
USK ENERGY / SEARCH

Search

Solutions, specialist directions and guides