3000TA-XL ppm Oxygen Analyser, Micro-Fuel Cell, 0–1 ppm Lowest Span up to 25 % O2

Electrochemical analyser for parts-per-million oxygen in nitrogen, argon and other high-purity gases. Its lowest span is 0–1 ppm, and the same instrument can read up to 25 % O2.

Measuring principle
micro-fuel cell
Spans
3 selectable, 0–1 ppm up to 0–10,000 ppm; 1–25 %
Bottom span offset
< 0.2 ppm
Alarm set points
0.00–10,000.00 ppm

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

Purity of a gas supply often hinges on a few ppm of oxygen, and this is where the 3000TA-XL earns its place: semiconductor fabs, high-purity gas distribution, air separation and liquefaction plants, inert blanketing, chemical reactors and petrochemical units.

The customer picks three spans. The smallest is 0–1 ppm full scale, ppm spans run up to 0–10,000 ppm, and a 1–25 % span covers percent levels. Offset on the bottom span stays below 0.2 ppm.

The sensing element is a micro-fuel cell, either the B-2CXL or the faster-recovering Insta-Trace B2CXL. An 8-bit microprocessor drives a two-line vacuum fluorescent display alongside an LED meter.

Gas should flow at 0.2 to 2.4 l/min, nominally 1.5 l/min, at 0.2 to 2.8 bar. For signals there is 0–1 V, an isolated 4–20 mA loop, a range ID signal and an RS-232 port. Two concentration alarms are adjustable between 0.00 and 10,000.00 ppm, and there is a system failure relay. Mains input is 85–250 V AC, 47–63 Hz.

Versions: -C adds separate zero and span gas connections with built-in valves; -CV has calibration valves; -V and -VS draw sample through the cell block by suction downstream instead of pushing it from upstream; -I is web-enabled with network I/O; -EU is fused for European mains.

The analyser is meant for indoor, non-hazardous locations, with a 19-inch rack version available. Conformity is declared to EN61326-1, EN55011 Class A Group 1, EN61010-1 and directive 2014/30/EU.

Technical specifications

Measuring principlemicro-fuel cell
Spans3 selectable, 0–1 ppm up to 0–10,000 ppm; 1–25 %
Bottom span offset< 0.2 ppm
Alarm set points0.00–10,000.00 ppm
Displaytwo-line VFD, LED meter
Analogue outputs0–1 V, isolated 4–20 mA, range ID
InterfaceRS-232; network on -I version
Alarms2 concentration, system failure
Gas flow0.2–2.4 l/min, nominal 1.5 l/min
Gas pressure0.2–2.8 bar
Supply85–250 V AC, 47–63 Hz
Locationindoor, non-hazardous; 19-inch rack

Request a price and lead time

3000TA-XL ppm Oxygen Analyser, Micro-Fuel Cell, 0–1 ppm Lowest Span up to 25 % O2. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

What is the lowest span on the 3000TA-XL?

0–1 ppm full scale, with offset on that span below 0.2 ppm. The customer chooses three spans in total.

Can it sample from a line below atmospheric pressure?

Yes. The -V and -VS versions pull gas through the cell block by suction behind it rather than pressure in front of it.

Which outputs are available to the control system?

0–1 V, isolated 4–20 mA, a range ID signal and RS-232; the -I version adds network I/O with a web interface.