3000TA Trace Oxygen Analyser, Micro-Fuel Cell Sensor, Lowest Scale 0–10 ppm
A panel-mounted instrument for measuring oxygen at ppm level in process and high-purity gases. Users set three ranges, the lowest being 0–10 ppm full scale, and a separate 0–25% range is kept for calibration.
- Measured component
- trace O2 in gas
- Sensor
- Insta Trace micro-fuel cell
- Lowest range
- 0–10 ppm O2 FS
- Calibration range
- 0–25% O2
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
At the heart of the 3000TA sits an Insta Trace micro-fuel cell, an electrochemical sensor chosen because it recovers quickly after being exposed to air. Factory default scales are LO 100 ppm, MED 1000 ppm and HI 10,000 ppm, and any of them can be reset, the narrowest permitted span being 0–10 ppm. Under stable temperature the error stays within ±1% FS, and the instrument resolves changes of 0.5% FS.
Speed depends on the range. At 25 °C the reading reaches 90% of full scale in under 10 s on scales of 1000 ppm and above, and in under 60 s on 0–10 ppm.
Typical duties include air separation and liquefaction plants, pure hydrocarbon streams, semiconductor fabs, gas purity certification and the blanketing of feedstocks and flammable liquids with inert gas. It also watches oxygen in gaseous monomers such as vinyl chloride, propylene, butadiene, isoprene and ethylene.
Signals leave the unit as 0–1 V dc plus an isolated 4–20 mA loop, alongside an RS-232 port. There are two concentration alarms and one system alarm, with relay contacts rated 3 A at 250 V ac. The mains input accepts 85–230 V ac at 50/60 Hz, and consumption stays within 20 VA.
The metal case is fitted flush into a panel and measures 8.70 × 6.96 × 12.2 in; sample lines connect with 1/4 in or 6 mm fittings. Ambient range is 0–50 °C. Installation is limited to indoor general-purpose areas, so it is not suited to hazardous locations.
Options include the 3000TA-C, which adds zero and span gas ports with built-in valves for automatic calibration, the high-sensitivity 3000TA-XL with a B-2CXL cell giving a 0–1 ppm span (±10 ppb when zeroed, ambient 5–35 °C), and the 3000TA-EU version.
Technical specifications
| Measured component | trace O2 in gas |
|---|---|
| Sensor | Insta Trace micro-fuel cell |
| Lowest range | 0–10 ppm O2 FS |
| Calibration range | 0–25% O2 |
| Accuracy | ±1% FS, stable temperature |
| Sensitivity | 0.5% FS |
| 90% response at 25 °C | <10 s (1000 ppm and above); <60 s (0–10 ppm) |
| Operating temperature | 0–50 °C |
| Analogue outputs | 0–1 V dc; isolated 4–20 mA |
| Interface | RS-232C |
| Power supply | 85–230 V ac, 50/60 Hz |
| Power consumption | 20 VA |
| Sample connections | 1/4 in or 6 mm |
| Mounting | flush panel |
| Case dimensions (W × H × D) | 8.70 × 6.96 × 12.2 in |
Request a price and lead time
3000TA Trace Oxygen Analyser, Micro-Fuel Cell Sensor, Lowest Scale 0–10 ppm. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
How narrow can the measuring span be set?
The standard 3000TA stops at a 0–10 ppm span. If you need 0–1 ppm, specify the 3000TA-XL with the B-2CXL sensor.
Can the analyser be installed in a hazardous area?
No. It is intended for indoor general-purpose areas only, so a hazardous location needs a different configuration.
What does the -C suffix add?
The 3000TA-C has separate zero and span gas ports and built-in valves driven by the electronics, which lets calibration run automatically.