International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

3000RS oxygen analyser, micro-fuel cell, ranges from 0–1 ppm to 0–100 % O2

A fixed oxygen analyser family that covers both ppm-level and percent-level O2 with one electrochemical sensing principle. Variants differ by range, and auto-calibration versions carry their own zero and span valves.

Sensor
micro-fuel cell
Ranges, 3000RS-P
0–1 % up to 0–100 %
Ranges, 3000RS-T
0–10 ppm up to 0–10,000 ppm
Range, -XL
from 0–1 ppm

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

Choosing a variant starts with the concentration to be measured. The 3000RS-P handles percent oxygen, 0–1 % up to 0–100 %, and answers in less than 10 s. For trace work the 3000RS-T spans 0–10 ppm up to 0–10,000 ppm; it responds in under 60 s on the 0–10 ppm range and under 10 s on 0–1,000 ppm. The -XL build reaches down to 0–1 ppm, with a response below 100 s. Add the -C suffix and the analyser gains dedicated zero and span gas ports plus internal control valves, so calibration can run unattended.

Typical duties include inert gas blanketing, air separation and liquefaction plants, semiconductor fabs, petrochemical process control and chemical reaction monitoring, as well as laboratory, quality assurance and gas certification work. The unit is intended for indoor, non-hazardous areas.

Oxygen is sensed by a micro-fuel cell with an expected life of around two years. Repeatability and linearity are both 1 % of full scale, the operating range is 0 to 50 °C and sample flow is 0.5 to 1.0 SLPM, watched by a digital flow meter. All gas connections are 1/8 in compression fittings, and 6 mm adapters suit metric tubing.

On the front sits a 3.5 in colour touchscreen LCD. Concentration is sent out as isolated 4–20 mA or 0–1 V, each paired with a range identification signal, and Profibus-DP and RS-232 links are fitted. Two adjustable concentration alarms and a system fault alarm switch Form-C relay contacts. The analyser runs from 24 VDC, or from 100–240 VAC 50/60 Hz mains via an adapter, drawing about 10 W, and carries CE marking with conformity to IEC 61010-1 and IEC 61326-1.

Technical specifications

Sensormicro-fuel cell
Ranges, 3000RS-P0–1 % up to 0–100 %
Ranges, 3000RS-T0–10 ppm up to 0–10,000 ppm
Range, -XLfrom 0–1 ppm
Response time< 10 s (P); < 60 s on 0–10 ppm (T); < 100 s on 0–1 ppm (-XL)
Repeatability / linearity1 % FS
Operating temperature0–50 °C
Sample flow0.5–1.0 SLPM
Display3.5 in colour touchscreen LCD
Analogue outputs4–20 mA (isolated) or 0–1 V, plus range ID
Alarms2 concentration + system fault, Form-C relays
CommunicationsProfibus-DP, RS-232
Power24 VDC; 100–240 VAC 50/60 Hz via adapter; 10 W
Gas connections1/8 in compression, 6 mm adapters
ConformityCE, IEC 61010-1, IEC 61326-1

Request a price and lead time

3000RS oxygen analyser, micro-fuel cell, ranges from 0–1 ppm to 0–100 % O2. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Which variant suits ppm-level purity checks on nitrogen?

The 3000RS-T covers 0–10 ppm up to 0–10,000 ppm. If a 0–1 ppm range is needed, specify the -XL version.

Can calibration be automated?

Yes. Versions with the -C suffix have separate zero and span gas ports and built-in valves for automatic calibration.

May the analyser be installed in a hazardous area?

No, it is intended for indoor, non-hazardous locations only.