3000PA panel-mount percent oxygen analyser, three ranges from 0–1 % to 0–100 %

A flush panel-mounted analyser for percent-level oxygen in process gases, using a Class B-1 micro-fuel cell. Option suffixes add auto-calibration, a 4–20 mA output, 316 stainless wetted parts or vacuum sampling.

Sensor
micro-fuel cell, Class B-1
Ranges
3 user-defined, from 0–1 % to 0–100 %
Calibration range
0–25 %
Accuracy
±2 % FS at constant temperature; ±5 % FS over operating range

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

Ordering is done by suffix. The 3000PA-C brings integral control valves with separate zero and span ports so calibration runs automatically, and the 3000PA-M adds an isolated 4–20 mA concentration output with a matching range ID signal. Where the gas attacks ordinary materials, the 3000PA-S has every wetted part in 316 stainless steel, while the 3000PA-V pulls the sample through with vacuum downstream of the cell block.

Three ranges are set by the user anywhere between 0–1 % and 0–100 %, with calibration carried out on 0–25 %. Accuracy is ±2 % of full scale at constant temperature and ±5 % of full scale across the operating band once the instrument has reached thermal equilibrium. At 25 °C the reading reaches 90 % of full scale within 10 s, and the unit works from 0 to 50 °C.

Typical gases include air separation and liquefaction streams, pure gaseous hydrocarbons, inert blanketing over feedstocks and flammable liquids, and monomers such as vinyl chloride, propylene, butadiene, isoprene and ethylene. It is also fitted for emissions and chemical reaction monitoring, semiconductor production, quality assurance and gas certification.

The case, 8.70 × 6.96 × 12.2 in, sits flush in a panel in a general-purpose area and runs from 85–250 VAC mains at 47–63 Hz. Sample enters at 3 to 40 psig through 1/4 in or 6 mm fittings, past a flow meter scaled 0.2–2.4 SLPM. A two-line vacuum fluorescent display sits next to an LED meter reading 0–100 %. Standard signals are 0–1 V dc for concentration and 0–1 V dc for range ID, plus an RS-232 port. Two adjustable concentration alarms and a system failure alarm drive relays rated up to 3 A at 250 V ac resistive, and remote calibration inputs accept 0 or 24 V dc. The analyser is CE marked to EN 61326-1 and EN 61010-1.

Technical specifications

Sensormicro-fuel cell, Class B-1
Ranges3 user-defined, from 0–1 % to 0–100 %
Calibration range0–25 %
Accuracy±2 % FS at constant temperature; ±5 % FS over operating range
Response90 % of FS in 10 s at 25 °C
Operating temperature0–50 °C
Outputs0–1 V dc concentration, 0–1 V dc range ID; 4–20 mA optional (-M)
Alarms2 concentration, 1 system failure; relays up to 3 A at 250 V ac
CommunicationsRS-232
Sample pressure3–40 psig; vacuum version (-V)
Flow meter0.2–2.4 SLPM
Gas connections1/4 in or 6 mm
Power85–250 VAC, 47–63 Hz
Mountingflush panel, general-purpose area
ConformityCE, EN 61326-1, EN 61010-1

Request a price and lead time

3000PA panel-mount percent oxygen analyser, three ranges from 0–1 % to 0–100 %. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Can the 3000PA measure ppm-level oxygen?

No, it is a percent analyser with a lowest range of 0–1 % O2. Three user-defined ranges extend up to 0–100 %.

Which options suit corrosive or sub-atmospheric samples?

The 3000PA-S has 316 stainless steel wetted parts for aggressive gases. The 3000PA-V draws the sample with vacuum downstream of the cell.

Does it offer a 4–20 mA output?

Yes, the 3000PA-M version adds an isolated 4–20 mA concentration output and a separate range ID output.