3000MA Panel-Mount Paramagnetic Percent Oxygen Analyser, 0-5 % to 0-100 %
A microprocessor-based percent oxygen analyser built around a paramagnetic cell and intended for flush panel mounting in general purpose areas. It was conceived with air separation plants in mind and reads oxygen in a wide variety of background gases.
- Measuring principle
- paramagnetic
- Ranges
- 3 user-set, 0-5 % to 0-100 %
- Calibration range
- 0-25 %
- Accuracy
- ±1 % FS at steady temperature; ±5 % FS over operating range
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
Air separation is the home ground of this analyser, but it is equally at home monitoring inert gas blanketing, chemical reactions and emissions, controlling petrochemical processes and checking or certifying gas mixtures. Ranges typically run from 0-5 % to 0-100 % O2, and lower spans down to 0-1 % can be supplied.
Three ranges are configured by the user, with autoranging between them. Because the response is linear on every range, a calibration made on one range holds when the instrument switches to another; calibration is carried out at up to 0-25 %. Once the analyser has reached thermal equilibrium, accuracy is ±1 % FS at a steady ambient and ±5 % FS across the full operating temperature band. Repeatability is 0.1 % FS, linearity 1 % O2, and both zero and span drift are 1 % per month. At 25 °C the reading gets to 90 % of its final value in less than 10 s.
Oxygen concentration appears on a 5 digit red LED readout, while a 2 line, 20 character vacuum fluorescent display carries the menus. Isolated analogue outputs report concentration and range ID, a contact signals system failure and two concentration alarms have fully programmable setpoints. The bidirectional RS-232C port lets the operator monitor the unit remotely and trigger zero and span calibration. Variants include the 3000MA-C with dedicated zero and span gas inlets and internal valves, the 3000MA-M on which the 4-20 mA current loop is live, and the 3000MA-EU supplied with spares for European use.
Sample gas should arrive at 5 to 20 psig, 1 to 2 SCFH and 0 °C to 50 °C, without condensation, through 1/4 inch or 6 mm fittings. Wetted parts are glass, steel, gold, Viton, acrylic glass and epoxy resin. The unit accepts 100-240 V AC, 50/60 Hz at 70 W maximum, operates at 0 °C to 45 °C, and complies with EN 61326-1 and EN 61010-1. The control unit is 6.96 inches high, 8.7 inches wide and 12.2 inches long.
Technical specifications
| Measuring principle | paramagnetic |
|---|---|
| Ranges | 3 user-set, 0-5 % to 0-100 % |
| Calibration range | 0-25 % |
| Accuracy | ±1 % FS at steady temperature; ±5 % FS over operating range |
| Repeatability | 0.1 % FS |
| Response time | 90 % FS in under 10 s at 25 °C |
| Displays | 5 digit LED; 2 x 20 character VFD |
| Alarms | system failure + 2 programmable setpoints |
| Interface | RS-232C |
| Supply | 100-240 V AC, 50/60 Hz, 70 W |
| Sample pressure | 5 to 20 psig |
| Sample flow | 1 to 2 SCFH |
| Operating temperature | 0 °C to 45 °C |
| Mounting | flush panel, general purpose area |
Request a price and lead time
3000MA Panel-Mount Paramagnetic Percent Oxygen Analyser, 0-5 % to 0-100 %. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
Does it need recalibrating when the range changes?
No. The response is linear on all three ranges, so one calibration covers them all; calibration can be done at up to 0-25 % O2.
What do the -C and -M suffixes add?
The 3000MA-C has dedicated zero and span gas inlets with internal valves, and on the 3000MA-M the 4-20 mA current loop output is live.
Can it be installed in a hazardous area?
No. Its area classification is general purpose, and it is designed for flush mounting in a control panel.