2120XL Plasma Emission Analyser for Trace Nitrogen in Argon or Helium

A 3U, 19-inch rack instrument that reads parts-per-billion levels of nitrogen in argon or helium by plasma ionisation and emission spectroscopy. It needs no column, carrier gas or switching valves, and its sensing element is not consumed.

Method
plasma ionisation, emission spectroscopy
Background gas
argon or helium
Ranges
0–100, 0–10, 0–1 ppm, fixed or auto
Detection limit
10 ppb

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

Nitrogen creeping into an argon or helium supply is usually the first sign that a seal, purifier or fill has gone wrong, so continuous measurement at trace level matters. Users include air separation plants, cryogenic tanker filling bays, specialty gas laboratories, argon purifiers, steelworks, chemical sites, welding gas systems, helium liquefiers and semiconductor fabs, as well as quality checks on tanker and cylinder fills.

The plasma emission detector runs under duty-cycle control, which limits contamination and coating inside the plasma cell. With no routine replacement intervals, upkeep is light, and internal parts are easy to reach when work is needed. Sample flow is set digitally anywhere in the 20–200 sccm band, inlet pressure may sit between 4 and 30 psig without upsetting the reading, and connections are 1/8 inch tube fittings.

The operator can choose 0–100, 0–10 or 0–1 ppm, fixed or with automatic switching, and higher ranges are optional; the detection limit of 10 ppb sits well below the smallest span. Resolution is 0.01 ppm under 10.0 ppm and 0.1 ppm from 10.0 ppm upwards. Accuracy and drift are each within 1% of full scale. Zero calibration uses purified argon or helium, span calibration uses ppm nitrogen in the same background gas.

Signals include an isolated 4–20 mA output, bidirectional RS-232, range identification by Form-A contacts and a 0–1 V DC signal, two programmable concentration alarms and a Form-C system alarm. Options cover an integral auto-calibration manifold, Profibus DP via the matching valve box, a moisture trap and a zero-gas scrubber. Supply is 110–240 V AC, 50/60 Hz, 60 W maximum; ambient range is 10°C to 35°C, the case measures 132.6 x 482.6 x 467.4 mm and weighs approximately 35 pounds.

Technical specifications

Methodplasma ionisation, emission spectroscopy
Background gasargon or helium
Ranges0–100, 0–10, 0–1 ppm, fixed or auto
Detection limit10 ppb
Resolution0.01 ppm below 10.0 ppm; 0.1 ppm from 10.0 ppm
Accuracy / drift1% FS each
Sample flow20–200 sccm, programmable
Inlet pressure4–30 psig
Gas connections1/8 inch tube fitting
Outputsisolated 4–20 mA, RS-232, 2 Form-C alarms
Power110–240 V AC, 50/60 Hz, 60 W max
Ambient temperature10°C to 35°C
Dimensions132.6 x 482.6 x 467.4 mm (3U)
Weightapproximately 35 pounds

Request a price and lead time

2120XL Plasma Emission Analyser for Trace Nitrogen in Argon or Helium. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

Which background gases can it handle?

Argon or helium. Flammable or toxic gases must never be introduced into the instrument.

Does it need a carrier gas or column?

No. The plasma detector works without switching valves, a column or carrier gas, and the sensing element is not consumed.

How does it connect to a control system?

Through the isolated 4–20 mA output, RS-232, Form-A and Form-C contacts, with Profibus DP available as an option via the valve box.