International supply. Engineering expertise.
office@usk.ua09:00–18:00 · Kyiv

TiR29 handheld thermal imager, 280 × 210 microbolometer, -20 to 150 °C range

A one-handed thermal imager for spotting temperature anomalies quickly on surveys and inspection rounds. Its dew point colour alarm marks surfaces cold enough for condensation, and fully radiometric images go into SmartView® for reporting.

Detector
uncooled microbolometer, 280 × 210
Spectral band
7.5–14 μm
Measurement range
-20 to 150 °C
Accuracy
±2 °C or 2 %, whichever is greater

Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.

Description

The 280 × 210 uncooled microbolometer covers the 7.5 to 14 μm band and measures from -20 to 150 °C, though calibration does not extend below -10 °C. Thermal sensitivity is 0.045 °C or better at a 30 °C target (45 mK), and accuracy is ±2 °C or 2 % at 25 °C nominal, whichever is greater. Spatial resolution is 2.86 mRad, with focus down to 7.5 cm. Minimum span is 2.0 °C in manual and 3 °C in automatic mode, and emissivity, reflected background and transmission corrections can all be set on screen.

Controls are laid out for one hand, including the manual Smart Focus and the capture, review and save sequence. A 9.1 cm diagonal VGA display at 640 × 480 offers bright or automatic backlight. The 2.0 megapixel visible camera is parallax corrected against the infrared image, and both picture-in-picture and full-screen infrared modes have three blending levels. Images are saved as .bmp or .jpeg, or as radiometric .is2 files.

Two lithium-ion smart packs each run for more than four hours at 50 % screen brightness and show their charge on a five-segment LED gauge. A full charge takes 2.5 hours in the two-bay charger or inside the camera, from 110 to 220 V AC mains at 50/60 Hz, and a 12 V vehicle adapter is available. The imager sleeps after 5 minutes idle and switches off after 30.

The body measures 27.7 × 12.2 × 17.0 cm and weighs 1.05 kg, is sealed to IP54 and survives a 2 m drop. Vibration is rated at 0.03 g²/Hz (3.8 grms) to IEC 68-2-6 and shock at 25 g to IEC 68-2-29. Safety approvals cover CSA C22.2 No. 61010-1-04 and UL 61010-1, with EMC to EN 61326-1. Operating temperature is -10 to 50 °C.

Technical specifications

Detectoruncooled microbolometer, 280 × 210
Spectral band7.5–14 μm
Measurement range-20 to 150 °C
Accuracy±2 °C or 2 %, whichever is greater
Thermal sensitivity (NETD)≤ 0.045 °C at 30 °C (45 mK)
Spatial resolution (IFOV)2.86 mRad
Minimum focus distance7.5 cm
Visible-light camera2.0 MP
Display9.1 cm, 640 × 480
Batteries2 × Li-ion, over 4 h each
ProtectionIP54, 2 m drop
Operating temperature-10 to 50 °C
Dimensions27.7 × 12.2 × 17.0 cm
Weight1.05 kg

Request a price and lead time

TiR29 handheld thermal imager, 280 × 210 microbolometer, -20 to 150 °C range. We reply within one business day.

* — required fields

Or email sales@usk.ua or call 044 339-50-20.

Questions and answers

What temperature range does the TiR29 measure?

It reads from -20 to 150 °C, with calibration not extending below -10 °C.

What is the dew point colour alarm for?

It colours areas whose temperature approaches the user-set dew point, showing where condensation is likely to form.

How long does a battery last?

Each of the two packs runs for more than four hours at 50 % screen brightness and recharges fully in 2.5 hours.