Ti45FT IR FlexCam Thermal Imager with IR-Fusion, 160 x 120, -20 °C to 600 °C
An industrial thermal imager that pairs a 160 x 120 microbolometer with a 1280 x 1024 visible light camera and a 5 inch display. IR-Fusion merges both pictures, so every thermogram shows exactly where the hot spot sits.
- Detector
- 160 x 120 VOx uncooled microbolometer
- Spectral band
- 8 µm to 14 µm
- Thermal sensitivity
- ≤80 mK at 30 °C
- Frame rate
- 30 Hz/60 Hz
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
At the core is an uncooled vanadium oxide (VOx) focal plane array of 160 x 120 pixels, sensitive from 8 µm to 14 µm with NETD of 80 mK or better at 30 °C and a 30 Hz/60 Hz frame rate. The standard 20 mm lens gives a 23° x 17° field of view, 2.60 mrad IFOV and focuses down to 0.15 m; a 10.5 mm germanium wide-angle lens (42° x 32°) and a 54 mm telephoto (9° x 6°) are available. SmartFocus lets the operator adjust focus with one finger, and both channels have 2x digital zoom.
The calibrated span runs from -20 °C to 600 °C, split into a -20 °C to 350 °C range and a 250 °C to 600 °C range; the high-temperature option adds 500 °C to 1200 °C. Accuracy is ±2 °C or 2 %, whichever is greater, and emissivity is set from 0.1 to 1.0 in steps of 0.01. Isotherms, movable spots and boxes with min, max and average, automatic hot and cold point detection and colour alarms above or below a threshold are all on board.
Predictive and plant maintenance, quality control and process monitoring are the main uses, along with electrical and mechanical systems, utilities, roofing, building envelopes, moisture tracing and R&D work.
The visible camera supports full thermal, full visual, blended and picture-in-picture views, and a Class 2 laser dot appears on screen when blending. Images carry 14 bit measurement data, go onto a 1 GB CompactFlash card and export to JPEG, BMP, PNG, GIF or TIFF; SmartView handles analysis and reporting, and a composite video output supports NTSC or PAL.
A field-replaceable lithium-ion smart battery gives 3 hours of continuous use, or 2 hours with IR-Fusion running, and a 110/220 V AC mains adapter is supplied. The housing is IP54, weighs 1.95 kg and measures 162 x 262 x 101 mm. It operates from -10 °C to 50 °C and can be stored at -40 °C to 70 °C.
Technical specifications
| Detector | 160 x 120 VOx uncooled microbolometer |
|---|---|
| Spectral band | 8 µm to 14 µm |
| Thermal sensitivity | ≤80 mK at 30 °C |
| Frame rate | 30 Hz/60 Hz |
| Field of view (20 mm lens) | 23° x 17° |
| IFOV (20 mm lens) | 2.60 mrad |
| Optional lenses | 10.5 mm (42° x 32°), 54 mm (9° x 6°) |
| Calibrated range | -20 °C to 600 °C |
| High-temperature option | 500 °C to 1200 °C |
| Accuracy | ±2 °C or 2 % |
| Visible light camera | 1280 x 1024 pixels |
| Storage medium | CompactFlash, 1 GB |
| Battery operation | 3 h (2 h with IR-Fusion) |
| Enclosure | IP54 |
| Weight | 1.95 kg |
Request a price and lead time
Ti45FT IR FlexCam Thermal Imager with IR-Fusion, 160 x 120, -20 °C to 600 °C. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
What temperature span does it cover?
-20 °C to 600 °C as calibrated, in two ranges of -20 °C to 350 °C and 250 °C to 600 °C. The high-temperature option extends the top end to 1200 °C.
Which lenses can be fitted?
The standard 20 mm lens (23° x 17°), a 10.5 mm wide-angle (42° x 32°) and a 54 mm telephoto (9° x 6°).
How are images stored and processed?
On a CompactFlash card with 14 bit radiometric data, exportable as JPEG, BMP, PNG, GIF or TIFF, and analysed in the SmartView reporting software.