Ti27 Thermal Imager, 240 x 180 Microbolometer, -20°C to +600°C
A handheld long-wave imager for electrical and mechanical surveys on industrial and commercial sites. The Ti27 pairs a 240 x 180 uncooled microbolometer with a 2.0 megapixel visible-light camera, so every thermogram can be tied to the spot it was taken.
- Measurement range
- -20°C to +600°C
- Accuracy
- ±2°C or 2%
- Detector
- uncooled microbolometer, 240 x 180 pixels
- Thermal sensitivity (NETD)
- ≤ 0.05°C at 30°C (50 mK)
Price, availability and lead time on request. We help with analogue selection, customs clearance and delivery.
Description
Readings span -20°C to +600°C, although the instrument is not calibrated below -10°C, and accuracy is ±2°C or 2%, whichever is greater. Thermal sensitivity is ≤ 0.05°C at a 30°C target (50 mK) across the 7.5 µm to 14 µm band, and the image refreshes at 9 Hz or 60 Hz depending on the version.
The housing is rated IP54 and has been drop-tested from 2 m with the standard lens; vibration is covered by IEC 68-2-6 and a 25 g shock by IEC 68-2-29. It works between -10°C and +50°C at 10% to 95% relative humidity without condensation, which suits switchrooms, plant rooms and, according to the manufacturer, clean rooms as well.
Focus is manual, set with one hand, down to 7.5 cm. Infrared and visible frames are aligned automatically with parallax correction, and picture-in-picture offers three blending levels alongside a full-screen infrared view. Emissivity, reflected background temperature and transmission can all be corrected on screen, while a user-set high-temperature colour alarm flags hot joints, motors and bearings during a walk-round.
Each image can carry up to 60 seconds of voice annotation. Files export as .bmp or .jpeg, or as fully radiometric .is2 for analysis and reporting in the bundled SmartView® software.
Two lithium-ion smart packs each last four hours or more and recharge in 2.5 hours; the imager can also run from the mains through the supplied 110 V AC to 220 V AC, 50/60 Hz power supply. It weighs 1.05 kg with battery fitted, and a two-year calibration interval is recommended.
Technical specifications
| Measurement range | -20°C to +600°C |
|---|---|
| Accuracy | ±2°C or 2% |
| Detector | uncooled microbolometer, 240 x 180 pixels |
| Thermal sensitivity (NETD) | ≤ 0.05°C at 30°C (50 mK) |
| Spectral band | 7.5 µm to 14 µm |
| Refresh rate | 9 Hz or 60 Hz |
| Visible-light camera | 2.0 megapixel |
| Minimum focus distance | 7.5 cm |
| Display | 9.1 cm, VGA 640 x 480 |
| Ingress protection | IP54 |
| Drop test | 2 m |
| Operating temperature | -10°C to +50°C |
| Power | two Li-ion packs, 4+ h each |
| Dimensions | 27.7 x 12.2 x 17.0 cm |
| Weight with battery | 1.05 kg |
Request a price and lead time
Ti27 Thermal Imager, 240 x 180 Microbolometer, -20°C to +600°C. We reply within one business day.
Or email sales@usk.ua or call 044 339-50-20.
Questions and answers
What temperature range does the Ti27 cover?
From -20°C to +600°C, with no calibration below -10°C. Accuracy is ±2°C or 2%, whichever is greater.
Can it be used in a dusty plant room?
Yes. The IP54 housing resists dust and water spray from any direction and survives a 2 m drop with the standard lens.
Which file formats does it save?
Non-radiometric .bmp and .jpeg, or fully radiometric .is2 for analysis in the bundled SmartView® software.